Ti-Containing Mesh Implants Modified by Heterostructures Based on Noble Metal: Fabrication and Study of Biological Activity.
- 1Nikolaev Institute of Inorganic Chemistry, Siberian Branch, Russian Academy of Sciences, Ac. Lavrentiev Ave. 3, 630090 Novosibirsk, Russia.
- 2Meshalkin National Medical Research Center, Ministry of Public Health of the Russian Federation, Rechkunovskaya Str. 15, 630055 Novosibirsk, Russia.
Abstract
Ti-based mesh implants ("titanium silk" TiS and NiTi alloy) offer excellent tissue integration for oncologic surgery but lack intrinsic antibacterial activity. To address this, we fabricated Ag nanoparticles and Ag/M (M = Pt, Au) heterostructures on mesh surfaces via vapor-phase deposition, with noble metal sublayers functionalized by Ag nanofilms and nanoclusters. The modified implants were comprehensively characterized using SEM, AFM, XPS, and ICP-AES and evaluated through in vitro biocompatibility and antibacterial assays. The functional performance of the heterostructures is governed by two key phenomena: a chemical sublayer effect and a geometry-dependent substrate effect, which together dictate ion release kinetics. The Ag/Au interface sustains Ag⁺ release, whereas Ag/Pt drives rapid Ag⁺ dissolution. All heterostructures demonstrated effective antibacterial activity against planktonic S. aureus within 24 h. Key factors responsible for reduction in the more resilient P. aeruginosa on the Ag/M heterostructures were hypothesized. Cytotoxicity was dictated by the released ion cocktail: Ag/Pt/NiTi exhibited elevated toxicity which may be associated with the combined release of Ag⁺, Ni²⁺ and Pt²⁺, whereas all Ag/Au-based heterostructures showed excellent biocompatibility (viability 75-108%). Notably, gold sublayers were associated with elevated IL-10 secretion by 72 h. Based on overall biological performance, Ag/Au heterostructures offering sustained antibacterial activity against planktonic bacteria, low cytotoxicity, and a cytokine profile that may be favorable for long-term integration are promising candidates for mesh implants. Also, h-Ag/Pt/TiS showed an optimal balance between rapid antibacterial action and biocompatibility and therefore appears to be a suitable candidate for further in vivo investigation.
Related
- Facile Fabrication of Multilayer Chemical Filters Based on Modified Activated Carbon and Activated Carbon Fiber Nonwovens for Airborne Molecular Contaminant Removal in Semiconductor Manufacturing
- Fabrication and Performance of Thermally Triggered Active Fire-Extinguishing Patches Based on Perfluorohexanone Microcapsule-Modified Epoxy Resin
- Fabrication and Characterization of 3D-Printed Polylactide Composite Scaffolds with Modified Graphene
- Structural Characterization and Visible-Light Photocatalytic Performance of AgBr-Modified Zr-Containing Titanium Silicalite-1 (TS-1) Composites
- Eco-Friendly Modified Fabrication of poly (lactic- co-glycolic acid) (PLGA) Nanoparticles from a Novel Utilized date seed Extract with Potent Antibacterial Properties isolated from Osteomyelitis Patients.
BibTeX
@article{dorovskikh2026ticontaining,
title = {{Ti-Containing Mesh Implants Modified by Heterostructures Based on Noble Metal: Fabrication and Study of Biological Activity.}},
author = {Svetlana Dorovskikh and Evgeniia Vikulova and David Sergeevichev and Tatiana Guselnikova and Aleksey Kolodin and Anastasiya Fedorenko and Alexander Zheravin and Natalya Morozova},
journal = {Journal of functional biomaterials},
year = {2026},
month = oct,
doi = {10.3390/jfb17110547},
url = {https://europepmc.org/article/MED/43082994},
note = {CC BY},
}
RIS
TY - JOUR
TI - Ti-Containing Mesh Implants Modified by Heterostructures Based on Noble Metal: Fabrication and Study of Biological Activity.
AU - Svetlana Dorovskikh
AU - Evgeniia Vikulova
AU - David Sergeevichev
AU - Tatiana Guselnikova
AU - Aleksey Kolodin
AU - Anastasiya Fedorenko
AU - Alexander Zheravin
AU - Natalya Morozova
PY - 2026
DA - 2026/10/27
JO - Journal of functional biomaterials
DO - 10.3390/jfb17110547
UR - https://europepmc.org/article/MED/43082994
AB - Ti-based mesh implants ("titanium silk" TiS and NiTi alloy) offer excellent tissue integration for oncologic surgery but lack intrinsic antibacterial activity. To address this, we fabricated Ag nanoparticles and Ag/M (M = Pt, Au) heterostructures on mesh surfaces via vapor-phase deposition, with noble metal sublayers functionalized by Ag nanofilms and nanoclusters. The modified implants were comprehensively characterized using SEM, AFM, XPS, and ICP-AES and evaluated through in vitro biocompatibility and antibacterial assays. The functional performance of the heterostructures is governed by two key phenomena: a chemical sublayer effect and a geometry-dependent substrate effect, which together dictate ion release kinetics. The Ag/Au interface sustains Ag⁺ release, whereas Ag/Pt drives rapid Ag⁺ dissolution. All heterostructures demonstrated effective antibacterial activity against planktonic S. aureus within 24 h. Key factors responsible for reduction in the more resilient P. aeruginosa on the Ag/M heterostructures were hypothesized. Cytotoxicity was dictated by the released ion cocktail: Ag/Pt/NiTi exhibited elevated toxicity which may be associated with the combined release of Ag⁺, Ni²⁺ and Pt²⁺, whereas all Ag/Au-based heterostructures showed excellent biocompatibility (viability 75-108%). Notably, gold sublayers were associated with elevated IL-10 secretion by 72 h. Based on overall biological performance, Ag/Au heterostructures offering sustained antibacterial activity against planktonic bacteria, low cytotoxicity, and a cytokine profile that may be favorable for long-term integration are promising candidates for mesh implants. Also, h-Ag/Pt/TiS showed an optimal balance between rapid antibacterial action and biocompatibility and therefore appears to be a suitable candidate for further in vivo investigation.
SN - 2079-4983
ER -