Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses
- 1Department of Cardiovascular Surgery, National Hospital Organization Kanmon Medical Center, Shimonoseki, Yamaguchi, Japan
- 2Department of Mechanical Engineering, Graduate School of Sciences and Technology for Innovation, Yamaguchi University, Ube, Yamaguchi, Japan
Abstract
Objectives: The aim of this study was to compare the effects of balloon-expandable and self-expanding stents on vessel wall mechanics and hemodynamics in a patient-specific saccular abdominal aortic aneurysm model.
Methods: A patient-specific aneurysm model was constructed using computed tomography images, and balloon-expandable and self-expanding stent models were created. Aortic wall stress was evaluated by contact analysis. Fluid–structure interaction analyses were performed at 5 representative phases selected from the cardiac cycle. Mean intra-aneurysmal flow velocity, wall shear stress (WSS), pressure loss, and energy loss were compared.
Results: The balloon-expandable stent produced greater localized aortic wall stress than the self-expanding stent, whereas wall stress was more uniformly distributed with the self-expanding stent. Both stents reduced the mean intra-aneurysmal flow velocity, WSS, pressure loss, and energy loss compared with the no-stent model. However, greater reductions in WSS, pressure loss, and energy loss were noted with the balloon-expandable stent than with the self-expanding stent.
Conclusions: Balloon-expandable and self-expanding stents exhibited different effects on vessel wall mechanics and hemodynamics. The findings indicate the importance of evaluating both hemodynamic and mechanical effects when assessing stent–vessel interactions.
Related
- Balloon-Expandable Vs Self-Expanding Valves for Transcatheter Aortic Valve Implantation in Internally Stented Surgical Valves
- Intrabiliary Rupture of a Hepatic Hydatid Cyst Causing Mechanical Biliary Obstruction: A Case Report of Endoscopic Management With a Self-Expandable Metal Stent
- Early Post-transcatheter Aortic Valve Implantation (TAVI) Haemodynamics With SAPIEN Balloon-Expandable and Evolut Self-Expanding Valve Platforms: A Retrospective Single-Centre Cohort Study.
- Comparison of intraoperative hemodynamic changes and recovery outcomes between non-menopausal and postmenopausal women undergoing total abdominal hysterectomy under general anesthesia.
- Artificial intelligence-based surrogate modeling of hemodynamic indicators in abdominal aortic aneurysms using computational fluid dynamics
BibTeX
@article{murakami2026mechanical,
title = {{Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses}},
author = {Masanori Murakami and Fei Jiang and Tatsuo Yamashita and Shinsuke Mochizuki},
journal = {Annals of Vascular Diseases},
year = {2026},
month = oct,
doi = {10.3400/avd.oa.26-00052},
url = {https://doi.org/10.3400/avd.oa.26-00052},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Mechanical and Hemodynamic Effects of Balloon-Expandable and Self-Expanding Stents in a Saccular Abdominal Aortic Aneurysm Model: Fluid–Structure Interaction Analyses
AU - Masanori Murakami
AU - Fei Jiang
AU - Tatsuo Yamashita
AU - Shinsuke Mochizuki
PY - 2026
DA - 2026/10/23
JO - Annals of Vascular Diseases
DO - 10.3400/avd.oa.26-00052
UR - https://doi.org/10.3400/avd.oa.26-00052
AB - Objectives: The aim of this study was to compare the effects of balloon-expandable and self-expanding stents on vessel wall mechanics and hemodynamics in a patient-specific saccular abdominal aortic aneurysm model. Methods: A patient-specific aneurysm model was constructed using computed tomography images, and balloon-expandable and self-expanding stent models were created. Aortic wall stress was evaluated by contact analysis. Fluid–structure interaction analyses were performed at 5 representative phases selected from the cardiac cycle. Mean intra-aneurysmal flow velocity, wall shear stress (WSS), pressure loss, and energy loss were compared. Results: The balloon-expandable stent produced greater localized aortic wall stress than the self-expanding stent, whereas wall stress was more uniformly distributed with the self-expanding stent. Both stents reduced the mean intra-aneurysmal flow velocity, WSS, pressure loss, and energy loss compared with the no-stent model. However, greater reductions in WSS, pressure loss, and energy loss were noted with the balloon-expandable stent than with the self-expanding stent. Conclusions: Balloon-expandable and self-expanding stents exhibited different effects on vessel wall mechanics and hemodynamics. The findings indicate the importance of evaluating both hemodynamic and mechanical effects when assessing stent–vessel interactions.
SN - 1881-641X
SN - 1881-6428
ER -