End-to-end film-based characterization of gated delivery latencies in a spot-scanned synchrotron-based proton therapy system.

  1. 1Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota, USA.

Abstract

Background

Temporal mismatches between gating signals and beam activation-beam-on and beam-off latencies-in gated particle beam therapy can compromise dose integrity. Yet standardized measurement techniques are lacking, and system-specific characterization, particularly for synchrotron-based proton systems, remains limited.Purpose

To implement a novel strategy for precise end-to-end gating latency measurements of a spot-scanned proton beam using Gafchromic film.Methods

Gating latencies have been measured for a Hitachi ProBeat accelerator equipped with a Varian RPM gating system. EBT4 films were positioned within a stack of acrylic blocks on a QUASAR respiratory motion platform operated with a 1.5 cm amplitude and 12 bpm sinusoidal pattern. A stainless-steel ball bearing (BB), positioned atop the film, facilitated localization. During irradiation, the BB produced a dose shadow on the film that indicated the phantom position. A proton plan consisting of a single 85.3 MeV spot was delivered to isocenter during a single gate-on/off interval initiated near the 25% phase, to maximize phantom velocity. Fifty-one measurements were made with a variety of gating intervals. Beam-on latency and delivery time produced a separation between the BB-shadow and the centroid of the Gaussian proton spot. Spot duration information was extracted from accelerator log files. BB-shadow/spot separation is modeled analytically based on phantom motion enabling the direct calculation of the total beam-on and beam-off latencies. Accelerator latency was measured directly with an oscilloscope, enabling isolation of the relative contributions of the RPM gating system.Results

An average beam-on latency of 142±27 ms is observed for the RPM+ProBeat system for the first gate of a delivery, which includes a known 69 ms beam extraction judgment time for the first proton spot in a spill from the Hitachi synchrotron. The corresponding average beam-off latency is 87±30ms . The average gate-on and gate-off latencies of the RPM system are calculated as 72±27ms and 81±30ms , respectively. The associated average contributions of the accelerator to the beam-on latencies are measured as 69.3±0.2ms (first gate in new spill) and 0.53±0.07ms (subsequent gates in existing spill) and similarly 6±4ms for beam-off latency.Conclusion

A novel, high-precision method for gating-latency measurements has been implemented using film to characterize gating latencies in a spot-scanned particle beam. Commissioning and regular quality assurance methods to precisely characterize the temporal accuracy of a gating system ensure treatment accuracy and will become increasingly vital to consider for higher-frequency applications, such as cardiac gating or FLASH-capable systems.

Related

  1. Magnet-Free Proton Therapy with 4D Pencil Beam Delivery Optimisation
    Nair N von Muehlenen, Florentin Bieder, Ye Zhang +1
    arXivPreprintarXiv licenceFull textBiological & medical physicsarXiv
  2. Dynamic Modulated Arc Therapy (DMAT): A Time Aware, Modulation Steered Optimization Framework for Next Generation Radiotherapy Delivery
    Taoran Li, Esa Kuusela, Emmi Ruokokoski +10
    arXivPreprintCC BY-NC-ND 4.0Full textBiological & medical physicsarXiv
  3. Impact of robust optimization and beam configuration on the distribution stability of LET under delivery uncertainties in carbon‐ion therapy
    Wisawa Phongprapun, Mutsumi Tashiro
    Journal of Applied Clinical Medical PhysicsPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationPMC
  4. Magnetic fields in blazar jets derived from synchrotron self-absorption
    Hubing Xiao, Junhang Liu, Shaopu Wang +6
    Astronomy & AstrophysicsPeer-reviewedPublisher termsHigh-energy & particle physicsCrossref
  5. Continuum Spectral Properties of Synchrotron Maser Emission: Comparison with Spontaneous Synchrotron Emission
    Lijie 利杰 Zhang 张, Dejin 德金 Wu 吴, Ling 玲 Chen 陈 +1
    The Astrophysical JournalPeer-reviewedCC BY 4.0Full textHigh-energy & particle physicsCrossref
BibTeX
@article{stottler2026endtoend,
  title = {{End-to-end film-based characterization of gated delivery latencies in a spot-scanned synchrotron-based proton therapy system.}},
  author = {Z S Stottler and J J Kruse and A J Deisher},
  journal = {Journal of applied clinical medical physics},
  year = {2026},
  month = oct,
  doi = {10.1002/acm2.70826},
  url = {https://europepmc.org/article/MED/42905951},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - End-to-end film-based characterization of gated delivery latencies in a spot-scanned synchrotron-based proton therapy system.
AU  - Z S Stottler
AU  - J J Kruse
AU  - A J Deisher
PY  - 2026
DA  - 2026/10/30
JO  - Journal of applied clinical medical physics
DO  - 10.1002/acm2.70826
UR  - https://europepmc.org/article/MED/42905951
AB  - Background  Temporal mismatches between gating signals and beam activation-beam-on and beam-off latencies-in gated particle beam therapy can compromise dose integrity. Yet standardized measurement techniques are lacking, and system-specific characterization, particularly for synchrotron-based proton systems, remains limited.Purpose  To implement a novel strategy for precise end-to-end gating latency measurements of a spot-scanned proton beam using Gafchromic film.Methods  Gating latencies have been measured for a Hitachi ProBeat accelerator equipped with a Varian RPM gating system. EBT4 films were positioned within a stack of acrylic blocks on a QUASAR respiratory motion platform operated with a 1.5 cm amplitude and 12 bpm sinusoidal pattern. A stainless-steel ball bearing (BB), positioned atop the film, facilitated localization. During irradiation, the BB produced a dose shadow on the film that indicated the phantom position. A proton plan consisting of a single 85.3 MeV spot was delivered to isocenter during a single gate-on/off interval initiated near the 25% phase, to maximize phantom velocity. Fifty-one measurements were made with a variety of gating intervals. Beam-on latency and delivery time produced a separation between the BB-shadow and the centroid of the Gaussian proton spot. Spot duration information was extracted from accelerator log files. BB-shadow/spot separation is modeled analytically based on phantom motion enabling the direct calculation of the total beam-on and beam-off latencies. Accelerator latency was measured directly with an oscilloscope, enabling isolation of the relative contributions of the RPM gating system.Results  An average beam-on latency of 142±27 ms is observed for the RPM+ProBeat system for the first gate of a delivery, which includes a known 69 ms beam extraction judgment time for the first proton spot in a spill from the Hitachi synchrotron. The corresponding average beam-off latency is 87±30ms . The average gate-on and gate-off latencies of the RPM system are calculated as 72±27ms and 81±30ms , respectively. The associated average contributions of the accelerator to the beam-on latencies are measured as 69.3±0.2ms (first gate in new spill) and 0.53±0.07ms (subsequent gates in existing spill) and similarly 6±4ms for beam-off latency.Conclusion  A novel, high-precision method for gating-latency measurements has been implemented using film to characterize gating latencies in a spot-scanned particle beam. Commissioning and regular quality assurance methods to precisely characterize the temporal accuracy of a gating system ensure treatment accuracy and will become increasingly vital to consider for higher-frequency applications, such as cardiac gating or FLASH-capable systems.
SN  - 1526-9914
ER  -