Highly Repeatable Measurement and Mechanical Modeling of Axial Insertion and Separation Forces in Hemodialysis Luer Connectors.
- 1Department of Clinical Engineering, Faculty of Health Care Sciences, Jikei University of Health Care Sciences, Osaka, Japan.
- 2Department of Clinical Engineering, Hanwa Memorial Hospital, Osaka, Japan.
- 3Independent Researcher, Osaka, Japan.
Abstract
Purpose
Unintended Luer connector disconnection during hemodialysis can cause blood loss and venous air embolism and persists despite Luer-lock mechanisms. The mechanical relationship between insertion force (Fin ) and separation force (Fsp ) remains unclear.Patients and methods
Fin and Fsp were measured under dry, axial loading using an automated linear stage and a six-axis force sensor, and a mechanical model based on Coulomb's law of friction was derived. Five male bloodline connector types and seven female needle or cannula connector types, including check valve-equipped cannulas, were tested in 35 combinations. Three female specimens per type were reused across male connectors, yielding 2001 cycles with insertion forces below 50 N. Linear mixed-effects models accounted for repeated measurements within specimens.Results
Fsp increased approximately linearly with Fin across all combinations (per-combination slopes, 0.57-0.99). Without a check valve, the pooled slope was 0.855 (95% confidence interval, 0.817-0.894); however, Fsp exceeded Fin in 20% of cycles, primarily at low insertion forces. With a check valve, the relationship had a negative intercept (Fsp = 0.758 Fin - 4.07 N), and certain combinations showed unstable engagement even at insertion forces near 10 N. The model yielded Fsp /Fin = tan(ρ₀ - θ)/tan(ρ' + θ), where ρ' is the insertion friction angle, ρ₀ denotes the static friction angle during engagement, and θ is the Luer taper angle. This ratio is below 1 when friction remains constant. The observed slopes corresponded to static friction coefficients of approximately 0.1-0.4, comparable to reported values for plastics.Conclusion
Clinical disconnection risk was not directly quantified. Under controlled dry, axial loading, Luer connectors generally separated at lower forces than the preceding insertion force. Axial retention differed among connector combinations and was lower in check valve-equipped cannulas. Evaluating connection stability may require consideration of mechanical engagement characteristics alongside dimensional standards.
Related
- Investigating the Relationship Between Mitral Leaflet Separation Index and Severity of Mitral Valve Stenosis Using Two and Three-Dimensional Measurement Methods in Trans Esophageal Echocardiography.
- Reduction of Klebsiella pneumoniae bacterial load after active disinfection of venous catheter connectors
- Repeatable individual variation in body temperature is associated with cold tolerance and exercise-induced hyperthermia in zebra finches
- Endovascular management of vascular complications in hemodialysis patients: a single-center experience
- Symptomatic Left Brachiocephlic Vein Occlusion in a Hemodialysis Patient
BibTeX
@article{matsui2026highly,
title = {{Highly Repeatable Measurement and Mechanical Modeling of Axial Insertion and Separation Forces in Hemodialysis Luer Connectors.}},
author = {Tomohiro Matsui and Atsushi Sakurai},
journal = {Medical devices (Auckland, N.Z.)},
year = {2026},
month = nov,
doi = {10.2147/mder.s644728},
url = {https://europepmc.org/article/MED/43081112},
note = {CC BY-NC},
}
RIS
TY - JOUR
TI - Highly Repeatable Measurement and Mechanical Modeling of Axial Insertion and Separation Forces in Hemodialysis Luer Connectors.
AU - Tomohiro Matsui
AU - Atsushi Sakurai
PY - 2026
DA - 2026/11/20
JO - Medical devices (Auckland, N.Z.)
DO - 10.2147/mder.s644728
UR - https://europepmc.org/article/MED/43081112
AB - Purpose Unintended Luer connector disconnection during hemodialysis can cause blood loss and venous air embolism and persists despite Luer-lock mechanisms. The mechanical relationship between insertion force (Fin ) and separation force (Fsp ) remains unclear.Patients and methods Fin and Fsp were measured under dry, axial loading using an automated linear stage and a six-axis force sensor, and a mechanical model based on Coulomb's law of friction was derived. Five male bloodline connector types and seven female needle or cannula connector types, including check valve-equipped cannulas, were tested in 35 combinations. Three female specimens per type were reused across male connectors, yielding 2001 cycles with insertion forces below 50 N. Linear mixed-effects models accounted for repeated measurements within specimens.Results Fsp increased approximately linearly with Fin across all combinations (per-combination slopes, 0.57-0.99). Without a check valve, the pooled slope was 0.855 (95% confidence interval, 0.817-0.894); however, Fsp exceeded Fin in 20% of cycles, primarily at low insertion forces. With a check valve, the relationship had a negative intercept (Fsp = 0.758 Fin - 4.07 N), and certain combinations showed unstable engagement even at insertion forces near 10 N. The model yielded Fsp /Fin = tan(ρ₀ - θ)/tan(ρ' + θ), where ρ' is the insertion friction angle, ρ₀ denotes the static friction angle during engagement, and θ is the Luer taper angle. This ratio is below 1 when friction remains constant. The observed slopes corresponded to static friction coefficients of approximately 0.1-0.4, comparable to reported values for plastics.Conclusion Clinical disconnection risk was not directly quantified. Under controlled dry, axial loading, Luer connectors generally separated at lower forces than the preceding insertion force. Axial retention differed among connector combinations and was lower in check valve-equipped cannulas. Evaluating connection stability may require consideration of mechanical engagement characteristics alongside dimensional standards.
SN - 1179-1470
ER -