The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System

  1. 1Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia
  2. 2Radiology and Medical Imaging Department, College of Applied Medical Sciences in Al-Kharj, Prince Sattam Bin Abdulaziz University, Al-Kharj 11942, Saudi Arabia

Abstract

Background: Ultrasound measurement of abdominal aortic aneurysm (AAA) diameter guides screening and surveillance, but whether students’ measurement competence on a conventional system transfers to an unfamiliar handheld device has rarely been assessed. Methods: Twenty-seven senior sonography students made three anteroposterior outer-to-outer measurements of two water-filled balloons embedded in gelatin (reference diameters 3.53 and 5.88 cm, calculated from the circumference before embedding), first on the familiar cart-based conventional system and then, in a fixed, non-randomised order, on a handheld system with which they had no prior hands-on experience. The primary outcomes were bias against the calculated reference, with equivalence testing (±5%), and the paired between-system difference, with Bland–Altman analysis; repeatability was assessed with repeatability coefficients. Results: Conventional measurements were equivalent to the calculated reference within ±5% (bias −0.09 cm, 95% CI −0.16 to −0.02; +0.07 cm, 95% CI −0.05 to +0.19). Handheld measurements were higher (+0.17 and +0.33 cm; both p < 0.001) and failed equivalence; the mean between-system difference was +0.26 cm (95% CI 0.18–0.34), without proportional bias. Single-reading repeatability coefficients were 0.31–0.67 cm. Conclusions: Students’ conventional-system measurements agreed closely with the calculated reference, although this reference was not verified after embedding. Because device type was confounded with operator inexperience and with a fixed testing order, which introduces an unquantifiable learning or fatigue effect, the 0.26 cm value is an observed between-system difference under the study conditions and cannot be attributed to the handheld device itself. Measurement competence did not transfer automatically to a new device, underlining the critical need for device-specific training before handheld systems are used for aortic measurement. These implications are educational and technical; phantom findings do not support clinical surveillance recommendations without confirmation in patients.

Related

  1. A lightweight portable ultrasound robot for autonomous liver sonography
    Zihan Li, Yixiao Xu, Lei Zhang +8
    Nature CommunicationsPeer-reviewedCC BY-NC-ND 4.0Full textGeneral medicineCrossref
  2. Repeatability of upper limb volume measurements based on the Kuhnke measurement protocol and frustum volume calculation in women after unilateral mastectomy: a cross-sectional methodological study
    Małgorzata Smoter, Robert Roczniok, Jarosław Muracki +3
    Scientific ReportsPeer-reviewedCC BY 4.0Full textOncologyCrossrefDOAJ
  3. A Competency Assessment Tool for Liver Contrast-Enhanced Ultrasound Among Sonography Trainees: Development and Multicenter Validation Study
    Yanping Ma, Bowen Zheng, Jie Ren +23
    JMIR Medical EducationPeer-reviewedCC BY 4.0Full textGeneral medicinePMC
  4. Comparison of Hockey-Stick and Conventional Linear Ultrasound Transducers for Detection of Non-Displaced Dentoalveolar Fractures: An Ex Vivo Paired Comparative Study
    Murat Mert Atapek, Sedef Ayşe Taşyapan, Elif Yıldızer
    TomographyPeer-reviewedCC BYFull textMedical imaging & radiologyEurope PMC
  5. Comparison of Atheroma Burden Changes Using Conventional vs. End-Diastolic Intravascular Ultrasound Segmentation ― Post Hoc Analysis of the PACMAN-AMI Study ―
    Nathan Angelo Lecaros Yap, Christos V. Bourantas, Sylvain Losdat +15
    Circulation JournalPeer-reviewedCC BY-NC-NDFull text2 citedSurgery & anaesthesiaJ-STAGE
BibTeX
@article{hassan2026need,
  title = {{The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System}},
  author = {Mahasin G. Hassan and Dema Alenazi and Salsabil Aldosari and Alanoud Alotaibi and Rema Alrajhi and Lamis Al-Sultan and Mohammed Alsaadi},
  journal = {Diagnostics},
  year = {2026},
  month = oct,
  doi = {10.3390/diagnostics16213524},
  url = {https://doi.org/10.3390/diagnostics16213524},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - The Need for Device-Specific Training Before Handheld Ultrasound Use: Agreement and Repeatability of Simulated Aortic Aneurysm Diameter Measurements by Senior Sonography Students on a Familiar Conventional and an Unfamiliar Handheld System
AU  - Mahasin G. Hassan
AU  - Dema Alenazi
AU  - Salsabil Aldosari
AU  - Alanoud Alotaibi
AU  - Rema Alrajhi
AU  - Lamis Al-Sultan
AU  - Mohammed Alsaadi
PY  - 2026
DA  - 2026/10/01
JO  - Diagnostics
DO  - 10.3390/diagnostics16213524
UR  - https://doi.org/10.3390/diagnostics16213524
AB  - Background: Ultrasound measurement of abdominal aortic aneurysm (AAA) diameter guides screening and surveillance, but whether students’ measurement competence on a conventional system transfers to an unfamiliar handheld device has rarely been assessed. Methods: Twenty-seven senior sonography students made three anteroposterior outer-to-outer measurements of two water-filled balloons embedded in gelatin (reference diameters 3.53 and 5.88 cm, calculated from the circumference before embedding), first on the familiar cart-based conventional system and then, in a fixed, non-randomised order, on a handheld system with which they had no prior hands-on experience. The primary outcomes were bias against the calculated reference, with equivalence testing (±5%), and the paired between-system difference, with Bland–Altman analysis; repeatability was assessed with repeatability coefficients. Results: Conventional measurements were equivalent to the calculated reference within ±5% (bias −0.09 cm, 95% CI −0.16 to −0.02; +0.07 cm, 95% CI −0.05 to +0.19). Handheld measurements were higher (+0.17 and +0.33 cm; both p < 0.001) and failed equivalence; the mean between-system difference was +0.26 cm (95% CI 0.18–0.34), without proportional bias. Single-reading repeatability coefficients were 0.31–0.67 cm. Conclusions: Students’ conventional-system measurements agreed closely with the calculated reference, although this reference was not verified after embedding. Because device type was confounded with operator inexperience and with a fixed testing order, which introduces an unquantifiable learning or fatigue effect, the 0.26 cm value is an observed between-system difference under the study conditions and cannot be attributed to the handheld device itself. Measurement competence did not transfer automatically to a new device, underlining the critical need for device-specific training before handheld systems are used for aortic measurement. These implications are educational and technical; phantom findings do not support clinical surveillance recommendations without confirmation in patients.
SN  - 2075-4418
ER  -