Feasibility of real-time 1.5-T MR-guided percutaneous irreversible electroporation in a porcine model
8 affiliations
- 1Department of Diagnostic and Interventional Radiology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany
- 2Research Campus STIMULATE, Otto-Hahn-Str. 2, 39106 Magdeburg, Germany
- 3Institute for Medical Engineering, Otto-von-Guericke-University Magdeburg, Universitätsplatz 2, 39106 Magdeburg, Germany
- 4Institute for Information Technology and Communications, Otto-von-Guericke-University Magdeburg, Universitätsplatz 2, 39106 Magdeburg, Germany
- 5Institute for Physics, Otto-von-Guericke-University Magdeburg, Universitätsplatz 2, 39106 Magdeburg, Germany
- 6Institute for Laboratory Animal Science and Central Animal Facility, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany
- 7Institute of Pathology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany
- 8Department of General, Visceral and Transplantation Surgery, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany
Abstract
This exploratory study assessed the procedural feasibility of in‑bore, real‑time magnetic resonance (MR)‑guided percutaneous irreversible electroporation (IRE) using a clinical 1.5-T system in a porcine model. Three healthy anesthetized juvenile swine underwent IRE in normal renal tissue using a prototype pulse generator and MR-conditional coaxial cannula as electrodes. One swine was ablated twice in the same kidney, resulting in statistically non-independent data. The procedure was guided by a balanced steady-state free precession sequence with orthogonal slices and intermittent position verifications using a radial three-dimensional gradient-echo sequence. Preprocedure radiofrequency noise and signal-to-noise ratio measurements showed slightly increased noise levels in generator standby mode without apparent imaging artifacts. Median procedure time was 48 min with an interquartile range of 17 min. Four hours post-procedure, the treatment zone appeared T1-hypointense in delayed contrast-enhanced imaging. For one treatment zone, hematoxylin and eosin staining (H&E), fluorescent terminal deoxynucleotidyl transferase-mediated dUTP nick-end labeling staining (TUNEL), and coregistration to MR imaging were performed. H&E showed minor hemorrhaging at the site of electrode insertion. TUNEL demonstrated a strong TUNEL-positive signal correlating with the T1-hypointense signal in delayed contrast-enhanced imaging, with a Dice-Sørensen coefficient of 67%. These findings demonstrate the procedural feasibility of MR-guided percutaneous IRE in clinical 1.5-T MR systems.
Key Points: Question Can MR-guided percutaneous irreversible electroporation be performed in a porcine model despite the lack of MR-compliant devices?
Findings In four irreversible electroporation procedures in three juvenile pigs, electrodes were successfully placed in the kidney using real-time guidance in a clinical 1.5-T scanner.
Relevance statement Real-time MR-guided percutaneous irreversible electroporation is procedurally feasible in a clinical 1.5-T MRI environment, while preserving image quality and enabling arbitrary slice positioning. Further research is required to confirm the reproducibility and enable the translation to oncological applications.
Graphical Abstract:
Related
- Irreversible electroporation for liver metastases not amenable to conventional treatments, an experience with technical challenges, clinical outcomes, and complications
- Histopathologic spectrum of focal liver lesions diagnosed by ultrasound-guided percutaneous liver biopsy and predictors of hepatocellular carcinoma: A single-center experience from Sub-Saharan Africa
- LNP-mediated BCL11A Editing Corrects Sickling Phenotypes and Preserves HSPC Fitness Compared to Electroporation
- Self-Guided Beta-Responsive Visual Neurofeedback in a Public Exhibition: Feasibility and Usability Randomized Controlled rial.
- Tension reduction from inter-membrane counter-deformation accelerates nuclear electroporation.
BibTeX
@article{belker2026feasibility,
title = {{Feasibility of real-time 1.5-T MR-guided percutaneous irreversible electroporation in a porcine model}},
author = {Othmar Belker and Simon Schröer and Thomas Gerlach and Max Joris Hubmann and Daniel Düx and Marcel Eisenmann and Christoph Fritsch and Julian Glandorf and Seben Sena Yerdelen and Inga Bruesch and Björn Hartleben and Sahar Arbabzadah and Jörg Mengwasser and Philipp Felgendreff and Georg Rose and Frank Wacker and Marcel Gutberlet and Bennet Hensen},
journal = {European Radiology Experimental},
year = {2026},
month = nov,
doi = {10.1186/s41747-026-00838-4},
url = {https://doi.org/10.1186/s41747-026-00838-4},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Feasibility of real-time 1.5-T MR-guided percutaneous irreversible electroporation in a porcine model
AU - Othmar Belker
AU - Simon Schröer
AU - Thomas Gerlach
AU - Max Joris Hubmann
AU - Daniel Düx
AU - Marcel Eisenmann
AU - Christoph Fritsch
AU - Julian Glandorf
AU - Seben Sena Yerdelen
AU - Inga Bruesch
AU - Björn Hartleben
AU - Sahar Arbabzadah
AU - Jörg Mengwasser
AU - Philipp Felgendreff
AU - Georg Rose
AU - Frank Wacker
AU - Marcel Gutberlet
AU - Bennet Hensen
PY - 2026
DA - 2026/11/03
JO - European Radiology Experimental
DO - 10.1186/s41747-026-00838-4
UR - https://doi.org/10.1186/s41747-026-00838-4
AB - This exploratory study assessed the procedural feasibility of in‑bore, real‑time magnetic resonance (MR)‑guided percutaneous irreversible electroporation (IRE) using a clinical 1.5-T system in a porcine model. Three healthy anesthetized juvenile swine underwent IRE in normal renal tissue using a prototype pulse generator and MR-conditional coaxial cannula as electrodes. One swine was ablated twice in the same kidney, resulting in statistically non-independent data. The procedure was guided by a balanced steady-state free precession sequence with orthogonal slices and intermittent position verifications using a radial three-dimensional gradient-echo sequence. Preprocedure radiofrequency noise and signal-to-noise ratio measurements showed slightly increased noise levels in generator standby mode without apparent imaging artifacts. Median procedure time was 48 min with an interquartile range of 17 min. Four hours post-procedure, the treatment zone appeared T1-hypointense in delayed contrast-enhanced imaging. For one treatment zone, hematoxylin and eosin staining (H&E), fluorescent terminal deoxynucleotidyl transferase-mediated dUTP nick-end labeling staining (TUNEL), and coregistration to MR imaging were performed. H&E showed minor hemorrhaging at the site of electrode insertion. TUNEL demonstrated a strong TUNEL-positive signal correlating with the T1-hypointense signal in delayed contrast-enhanced imaging, with a Dice-Sørensen coefficient of 67%. These findings demonstrate the procedural feasibility of MR-guided percutaneous IRE in clinical 1.5-T MR systems. Key Points: Question Can MR-guided percutaneous irreversible electroporation be performed in a porcine model despite the lack of MR-compliant devices? Findings In four irreversible electroporation procedures in three juvenile pigs, electrodes were successfully placed in the kidney using real-time guidance in a clinical 1.5-T scanner. Relevance statement Real-time MR-guided percutaneous irreversible electroporation is procedurally feasible in a clinical 1.5-T MRI environment, while preserving image quality and enabling arbitrary slice positioning. Further research is required to confirm the reproducibility and enable the translation to oncological applications. Graphical Abstract:
SN - 2509-9280
ER -