PhysiCal-PET: Physics-driven neural networks for the rapid position Calibration of PET Scanners
Abstract
Depth-of-interaction (DOI) information is essential for mitigating parallax error and preserving uniform spatial resolution in high-resolution and long-axial-field-of-view positron emission tomography (PET). Conventional DOI calibration relies on mechanically scanned collimated sources or detector-specific bench measurements, which become impractical as the number of detector modules grows. We present PhysiCal-PET, a physics-driven neural-network framework for fast, label-free DOI calibration of fully assembled PET scanners. Coincidences are acquired from an uncollimated point source placed at a few known positions, and a network mapping DOI-sensitive detector signals to interaction depth is trained by minimizing the distance between the known source position and the line of response (LOR) defined by the two predicted interaction positions, so that the annihilation geometry alone provides supervision, without measured DOI labels. In realistic simulations, the DOI mean absolute error decreased in parallel with the geometric loss and converged to 0.98 mm, confirming that the loss is a faithful surrogate for the unobservable DOI error. Applied to a preclinical scanner with 144 pixelated dual-ended-readout detector modules, the method calibrated the entire system from 11 manually placed source positions in 55 min of acquisition, reducing the trimmed source-to-LOR distance by 24%, improving the radial point-source FWHM from 4.99 to 2.90 mm (42%), and raising the Derenzo peak-to-valley ratio of the 1.7 mm rods from 1.28 to 1.88 while separating the 1.2 mm rods, which remain merged without calibration. Requiring no collimation, motion stages, or reference measurements, and scaling favorably with system sensitivity, PhysiCal-PET offers a practical route toward autonomous, routine DOI calibration of fully assembled PET scanners with many detectors.
Related
- Physics-Informed Self-Supervised Learning for Joint Wire Calibration and Interaction Position Reconstruction in Multi-Wire Parallel Plate Avalanche Counters
- Automated Physics-Informed Neural-Networks-Based Calibration of Highly Segmented Silicon Telescopes
- Impact of the position effects in the energy calibration of the BULLKID detector
- A Clarifying Note on the Position-Momentum Correspondence: Pontryagin Duality, Fourier Transport, and Physical Normalization
- Loss-Tolerant Quantum Position Verification for Metropolitan Area Networks
BibTeX
@misc{abushawish2026physicalpet,
title = {{PhysiCal-PET: Physics-driven neural networks for the rapid position Calibration of PET Scanners}},
author = {Mojahed Abushawish and Pablo Galve and Iván Colmenarejo Allavena and Carlos Galindo Morata and Sara Gaitan Dominguez and Jose Manuel Udias},
howpublished = {arXiv},
year = {2026},
month = oct,
doi = {10.48550/arxiv.2610.24193},
eprint = {2610.24193},
archivePrefix = {arXiv},
url = {https://arxiv.org/abs/2610.24193},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - UNPB
TI - PhysiCal-PET: Physics-driven neural networks for the rapid position Calibration of PET Scanners
AU - Mojahed Abushawish
AU - Pablo Galve
AU - Iván Colmenarejo Allavena
AU - Carlos Galindo Morata
AU - Sara Gaitan Dominguez
AU - Jose Manuel Udias
PY - 2026
DA - 2026/10/16
PB - arXiv
DO - 10.48550/arxiv.2610.24193
UR - https://arxiv.org/abs/2610.24193
AB - Depth-of-interaction (DOI) information is essential for mitigating parallax error and preserving uniform spatial resolution in high-resolution and long-axial-field-of-view positron emission tomography (PET). Conventional DOI calibration relies on mechanically scanned collimated sources or detector-specific bench measurements, which become impractical as the number of detector modules grows. We present PhysiCal-PET, a physics-driven neural-network framework for fast, label-free DOI calibration of fully assembled PET scanners. Coincidences are acquired from an uncollimated point source placed at a few known positions, and a network mapping DOI-sensitive detector signals to interaction depth is trained by minimizing the distance between the known source position and the line of response (LOR) defined by the two predicted interaction positions, so that the annihilation geometry alone provides supervision, without measured DOI labels. In realistic simulations, the DOI mean absolute error decreased in parallel with the geometric loss and converged to 0.98 mm, confirming that the loss is a faithful surrogate for the unobservable DOI error. Applied to a preclinical scanner with 144 pixelated dual-ended-readout detector modules, the method calibrated the entire system from 11 manually placed source positions in 55 min of acquisition, reducing the trimmed source-to-LOR distance by 24%, improving the radial point-source FWHM from 4.99 to 2.90 mm (42%), and raising the Derenzo peak-to-valley ratio of the 1.7 mm rods from 1.28 to 1.88 while separating the 1.2 mm rods, which remain merged without calibration. Requiring no collimation, motion stages, or reference measurements, and scaling favorably with system sensitivity, PhysiCal-PET offers a practical route toward autonomous, routine DOI calibration of fully assembled PET scanners with many detectors.
ER -