Skip to content
DraconiticResearch
Search Sources API
203,203 studies31 sourcesUpdated 40 min ago
PreprintarXiv licencearXiv

Cone-beam artifact reduction in Gamma Knife CBCT images using a line-arc-line scan trajectory

Abstract

Objective. Gamma Knife cone-beam computed tomography (CBCT) images suffer from distinct cone-beam artifacts for some patients, due to the conical X-ray beam which is oriented to intersect the detector perpendicularly at its inferior edge. The use of an exotic scan trajectory which provides more complete sampling across the field of view than the current arc can reduce cone-beam artifacts. In this study, the feasibility of a line-arc-line scan trajectory for the Gamma Knife CBCT is assessed through a proof of concept. Approach. Using a customized research Gamma Knife, a Catphan 503 and an anthropomorphic head phantom are imaged for different exotic scan trajectories, consisting of combinations of arcs and lines. The CBCT images for each trajectory are then reconstructed with an iterative algorithm. Main results. A line-arc-line trajectory provides the largest image quality improvement among the investigated trajectories, with no noticeable cone-beam artifacts in the CBCT images: the axial interfaces between modules of the Catphan are well defined, and the superior edge of the phantom on the axial axis is sharper by 92\%, compared with the current arc trajectory. For the head CBCT images, the proposed trajectory removes most of the cone-beam artifact on the superior side of the skull, characteristic of the current Gamma Knife CBCT images. The artifact reduction also leads to a visual improvement in terms of uniformity in both phantoms. Significance. A line-arc-line CBCT scan trajectory would be feasible on the Gamma Knife with limited changes to the current configuration, and could produce images with improved image quality.

Related

  1. Non-circular scan trajectories for reducing cone-beam artifacts in Gamma Knife CBCT images: a simulation study
    Alexandra Alain-Beaudoin, Håkan Nordström, Luc Beaulieu +1
    arXivPreprintCC BY 4.0Full textBiological & medical physicsarXiv
  2. Comparative Analysis of Commercial Independent Dose Calculation Systems for Gamma Knife Stereotactic Radiosurgery
    Michael D. Lowe, Jack C. Nelson, Joe Caron +4
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
  3. Dosimetric Comparison of Functional Radiosurgery Treatments for Trigeminal Neuralgia, Essential Tremor, and Obsessive-compulsive Disorder Using Gamma Knife, CyberKnife, and Linear Accelerators
    Carlos Daniel Venencia, Crystian Wilian Chagas Saraiva, Gabriela Reis dos Santos de Jesus
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
  4. Treatment plan quality comparison between ZAP-Axon® Krylov optimizer and Leksell Gamma Knife® Lightning optimizer for brain metastasis.
    Yongsook C Lee, Alexander Maslowski, Rupesh Kotecha +6
    Journal of applied clinical medical physicsPeer-reviewedCC BYFull textApplied physics & instrumentationEurope PMCPMC
  5. Image-Domain CT Metal Artifact Reduction Using an Endpoint-Conditioned Diffusion Bridge
    Hang Yu, Mojtaba Safari, Shansong Wang +2
    arXivPreprintCC BY 4.0Full textBiological & medical physicsarXivPMC
BibTeX
@misc{alainbeaudoin2026conebeam,
  title = {{Cone-beam artifact reduction in Gamma Knife CBCT images using a line-arc-line scan trajectory}},
  author = {Alexandra Alain-Beaudoin and Håkan Nordström and Luc Beaulieu and Joakim da Silva},
  howpublished = {arXiv},
  year = {2026},
  month = sep,
  doi = {10.48550/arxiv.2609.30169},
  eprint = {2609.30169},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2609.30169},
}
RIS
TY  - UNPB
TI  - Cone-beam artifact reduction in Gamma Knife CBCT images using a line-arc-line scan trajectory
AU  - Alexandra Alain-Beaudoin
AU  - Håkan Nordström
AU  - Luc Beaulieu
AU  - Joakim da Silva
PY  - 2026
DA  - 2026/09/24
PB  - arXiv
DO  - 10.48550/arxiv.2609.30169
UR  - https://arxiv.org/abs/2609.30169
AB  - Objective. Gamma Knife cone-beam computed tomography (CBCT) images suffer from distinct cone-beam artifacts for some patients, due to the conical X-ray beam which is oriented to intersect the detector perpendicularly at its inferior edge. The use of an exotic scan trajectory which provides more complete sampling across the field of view than the current arc can reduce cone-beam artifacts. In this study, the feasibility of a line-arc-line scan trajectory for the Gamma Knife CBCT is assessed through a proof of concept. Approach. Using a customized research Gamma Knife, a Catphan 503 and an anthropomorphic head phantom are imaged for different exotic scan trajectories, consisting of combinations of arcs and lines. The CBCT images for each trajectory are then reconstructed with an iterative algorithm. Main results. A line-arc-line trajectory provides the largest image quality improvement among the investigated trajectories, with no noticeable cone-beam artifacts in the CBCT images: the axial interfaces between modules of the Catphan are well defined, and the superior edge of the phantom on the axial axis is sharper by 92\%, compared with the current arc trajectory. For the head CBCT images, the proposed trajectory removes most of the cone-beam artifact on the superior side of the skull, characteristic of the current Gamma Knife CBCT images. The artifact reduction also leads to a visual improvement in terms of uniformity in both phantoms. Significance. A line-arc-line CBCT scan trajectory would be feasible on the Gamma Knife with limited changes to the current configuration, and could produce images with improved image quality.
ER  -