Agreement Between Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry (DXA) for the Magnitude and Direction of Between-Limb Lean Mass Asymmetry in Trail Runners.

4 affiliations
  1. 1Institut Nacional d'Educació Física de Catalunya (INEFC), Av. de l'Estadi 12-22, 08038 Barcelona, Spain.
  2. 2INEFC-Barcelona Research Group on Sport Sciences (GRCEIB), Av. de l'Estadi 12-22, 08038 Barcelona, Spain.
  3. 3Catalan School of Kinanthropometry (ECC-INEFC), Av. de l'Estadi 12-22, 08038 Barcelona, Spain.
  4. 4BCN Body Scan SL, Av. de l'Estadi 12-22, 08038 Barcelona, Spain.

Abstract

Background: Between-limb lean-mass asymmetry has been proposed for injury-risk and return-to-sport monitoring. Dual-energy X-ray absorptiometry (DXA), the reference method, requires a fixed installation and ionizing radiation; segmental bioelectrical impedance analysis (BIA) is fast, inexpensive and field-applicable. Agreement for the asymmetry itself, rather than absolute limb values, is unresolved. Methods: In this cross-sectional method-comparison study, 127 trail runners (90 men, 37 women) underwent eight-electrode segmental BIA and DXA in the same session. An asymmetry index, the between-limb difference relative to the limb mean, was computed from raw leg resistance (BIA) and leg lean mass (DXA), and assessed by Bland-Altman analysis, concordance correlation (CCC) and kappa. Results: Mean bias was -0.97 percentage points (pp), but the 95% limits of agreement spanned -6.67 to +4.73 pp, three to four times the observed interquartile range; CCC was 0.414. The same leg was identified in 70.9% of runners (kappa 0.309) against a right-side predominance of 75.6%, and the BIA classification was not significantly better than a constant right-leg rule (p = 0.451). Conclusions: Segmental BIA cannot replace DXA for assessing between-limb lean-mass asymmetry in individual trail runners. The disagreement between methods exceeds the asymmetries observed, and directional agreement must be read against the base rate.

Related

  1. Agreement of Bioelectrical Impedance Analysis Prediction Equations with Dual-Energy X-Ray Absorptiometry for Tracking Resistance Training-Induced Changes in Fat-Free Mass in Older Women.
    Alex S Ribeiro, André N S Cahebo, Amilton T Canguali +9
    Sports (Basel, Switzerland)Peer-reviewedCC BYFull textPhysiologyEurope PMC
  2. Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults
    Elodie Piche, Emeline Michel, Olivier Guerin +2
    SensorsPeer-reviewedCC BY 4.0Full textPhysiologyHAL
  3. Bioelectrical impedance analysis versus dual-energy X-ray absorptiometry for sarcopenia assessment in type 2 diabetes mellitus: A systematic review and meta-analysis
    Shobhit Raj, Sahana Shetty, Renuka Suvarna +2
    PLOS OnePeer-reviewedCC BY 4.0Full textGeneral biochemistryGeneral biologyPLOSPMC
  4. Short-Term Changes in Segmental Multi-Frequency Bioelectrical Impedance Measurements Following Water and Banana Intake: A Fixed-Order Within-Subject Study.
    Erika Čermáková, Monika Kunzová, Jana Olejníčková +2
    International journal of medical sciencesPeer-reviewedCC BYFull textPhysiologyEurope PMC
  5. Missing Link between Glycemic Control and Bioelectrical Impedance Analysis
    Akinori Higaki
    Internal MedicinePeer-reviewedCC BY-NC-NDFull textPhysiologyJ-STAGE
BibTeX
@article{pinolgranadino2026agreement,
  title = {{Agreement Between Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry (DXA) for the Magnitude and Direction of Between-Limb Lean Mass Asymmetry in Trail Runners.}},
  author = {Nil Piñol-Granadino and Neus Guinovart and Marta Carrasco-Marginet and Sílvia Puigarnau and Fabrizio Gravina-Cognetti and Quim Rosales and Alfredo Irurtia},
  journal = {Journal of functional morphology and kinesiology},
  year = {2026},
  month = nov,
  doi = {10.3390/jfmk11040452},
  url = {https://europepmc.org/article/MED/43081351},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Agreement Between Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry (DXA) for the Magnitude and Direction of Between-Limb Lean Mass Asymmetry in Trail Runners.
AU  - Nil Piñol-Granadino
AU  - Neus Guinovart
AU  - Marta Carrasco-Marginet
AU  - Sílvia Puigarnau
AU  - Fabrizio Gravina-Cognetti
AU  - Quim Rosales
AU  - Alfredo Irurtia
PY  - 2026
DA  - 2026/11/01
JO  - Journal of functional morphology and kinesiology
DO  - 10.3390/jfmk11040452
UR  - https://europepmc.org/article/MED/43081351
AB  - Background: Between-limb lean-mass asymmetry has been proposed for injury-risk and return-to-sport monitoring. Dual-energy X-ray absorptiometry (DXA), the reference method, requires a fixed installation and ionizing radiation; segmental bioelectrical impedance analysis (BIA) is fast, inexpensive and field-applicable. Agreement for the asymmetry itself, rather than absolute limb values, is unresolved. Methods: In this cross-sectional method-comparison study, 127 trail runners (90 men, 37 women) underwent eight-electrode segmental BIA and DXA in the same session. An asymmetry index, the between-limb difference relative to the limb mean, was computed from raw leg resistance (BIA) and leg lean mass (DXA), and assessed by Bland-Altman analysis, concordance correlation (CCC) and kappa. Results: Mean bias was -0.97 percentage points (pp), but the 95% limits of agreement spanned -6.67 to +4.73 pp, three to four times the observed interquartile range; CCC was 0.414. The same leg was identified in 70.9% of runners (kappa 0.309) against a right-side predominance of 75.6%, and the BIA classification was not significantly better than a constant right-leg rule (p = 0.451). Conclusions: Segmental BIA cannot replace DXA for assessing between-limb lean-mass asymmetry in individual trail runners. The disagreement between methods exceeds the asymmetries observed, and directional agreement must be read against the base rate.
SN  - 2411-5142
ER  -