Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems

  1. 1Medical Physics Department, Maria Skłodowska-Curie National Research Institute of Oncology, Warsaw, Poland

Abstract

Introduction: The Varian Ethos adaptive radiotherapy system was upgraded from version 1.0 to 2.0, introducing an enhanced multi-leaf collimator (MLC) dose calculation model and a new optimization engine. We evaluated whether this affected agreement between calculated and measured dose, and plan complexity.

Material and methods: Twenty-four plans (14 intensity-modulated radiotherapy [IMRT], 10 volumetric modulated arc therapy [VMAT]; 26 measurement configurations) covering 10 anatomical sites were analyzed in three matched configurations: original Ethos 1.0 plans (E1.0), the same plans recalculated in Ethos 2.0 with the new MLC model only (E1.0→2.0), and plans re-optimized in Ethos 2.0 (E2.0). Quality assurance comprised ionization chamber point-dose measurement and global gamma analysis on an ArcCHECK array at 3%/3 mm, 3%/2 mm and 2%/2 mm (10% threshold). Complexity was quantified with UCoMX v1.1; groups were compared with the Wilcoxon signed-rank test.

Results: Mean signed chamber deviation was −1.59 ± 0.54% (E1.0), −1.54 ± 0.57% (E1.0→2.0) and −1.39 ± 0.63% (E2.0), with no significant differences (p > 0.6); all met the ±3% criterion. Gamma pass rates at 3%/3 mm were 99.2 ± 0.9%, 98.7 ± 1.0% (p < 0.001) and 98.6 ± 2.1% (p = 0.059). Re-optimized plans used 12% fewer monitor units (1,169 ± 395 vs. 1,338 ± 471, p < 0.001) and were simpler (MCSv 0.31 to 0.35; SAS10 mm 0.27 to 0.20; both p < 0.001).

Conclusion: The upgrade preserved quality-assurance compliance and produced simpler, lower-monitor-unit plans, but the systematic −1.5% offset between calculated and measured point dose persisted, showing the new MLC model alone does not remove it.

Related

  1. Dosimetric and Treatment Delivery Efficiency Comparison of Two Ring-gantry Linear Accelerator Platforms for Liver Stereotactic Body Radiotherapy
    Srikanth Guduru, Rohith Singareddy, N. V. N. Madhusudhana Sresty +2
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
  2. Dosimetric Comparison of One-, Two-, Three-, and Four-Arc Volumetric Modulated Arc Therapy Plans in Cervical Carcinoma: Evaluation of Plan Quality, Organ-at-Risk Sparing, and Treatment Efficiency.
    Priyadharshini G, Himanshi Khattar, Piyush Kumar
    CureusPeer-reviewedCC BYFull textApplied physics & instrumentationEurope PMC
  3. Dosimetric Verification Using Tissue-Equivalent Phantoms
    Naayem Abdulraheem Salem Bakran, Ali Sid Ahmed Mohammed Ali, Al-Nazer Osman Mohammed Hamza +2
    Advances in Radiation OncologyPeer-reviewedCC BY-NC-ND 4.0Full textApplied physics & instrumentationPMC
  4. Impact of CT image co-registration on dosimetric verification in ¹²⁵I seed implantation for lung cancer
    Xu Ke, Yu Huimin, Zhang Hongtao +5
    BMC CancerPeer-reviewedCC BY-NC-ND 4.0Full textApplied physics & instrumentationPMC
  5. Dosimetric Comparison of Functional Radiosurgery Treatments for Trigeminal Neuralgia, Essential Tremor, and Obsessive-compulsive Disorder Using Gamma Knife, CyberKnife, and Linear Accelerators
    Carlos Daniel Venencia, Crystian Wilian Chagas Saraiva, Gabriela Reis dos Santos de Jesus
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
BibTeX
@article{wesolowska2026comparison,
  title = {{Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems}},
  author = {Paulina Wesołowska and Joanna Iwanowska-Hanke and Anna Paciorkiewicz and Sandra Maluszczak and Mikołaj Tarchalski and Bartłomiej Sadowski and Marta Fillmann and Agnieszka Walewska},
  journal = {Acta Oncologica},
  year = {2026},
  month = nov,
  doi = {10.2340/1651-226x.2026.46866},
  url = {https://doi.org/10.2340/1651-226x.2026.46866},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems
AU  - Paulina Wesołowska
AU  - Joanna Iwanowska-Hanke
AU  - Anna Paciorkiewicz
AU  - Sandra Maluszczak
AU  - Mikołaj Tarchalski
AU  - Bartłomiej Sadowski
AU  - Marta Fillmann
AU  - Agnieszka Walewska
PY  - 2026
DA  - 2026/11/18
JO  - Acta Oncologica
DO  - 10.2340/1651-226x.2026.46866
UR  - https://doi.org/10.2340/1651-226x.2026.46866
AB  - Introduction: The Varian Ethos adaptive radiotherapy system was upgraded from version 1.0 to 2.0, introducing an enhanced multi-leaf collimator (MLC) dose calculation model and a new optimization engine. We evaluated whether this affected agreement between calculated and measured dose, and plan complexity.  Material and methods: Twenty-four plans (14 intensity-modulated radiotherapy [IMRT], 10 volumetric modulated arc therapy [VMAT]; 26 measurement configurations) covering 10 anatomical sites were analyzed in three matched configurations: original Ethos 1.0 plans (E1.0), the same plans recalculated in Ethos 2.0 with the new MLC model only (E1.0→2.0), and plans re-optimized in Ethos 2.0 (E2.0). Quality assurance comprised ionization chamber point-dose measurement and global gamma analysis on an ArcCHECK array at 3%/3 mm, 3%/2 mm and 2%/2 mm (10% threshold). Complexity was quantified with UCoMX v1.1; groups were compared with the Wilcoxon signed-rank test.  Results: Mean signed chamber deviation was −1.59 ± 0.54% (E1.0), −1.54 ± 0.57% (E1.0→2.0) and −1.39 ± 0.63% (E2.0), with no significant differences (p > 0.6); all met the ±3% criterion. Gamma pass rates at 3%/3 mm were 99.2 ± 0.9%, 98.7 ± 1.0% (p < 0.001) and 98.6 ± 2.1% (p = 0.059). Re-optimized plans used 12% fewer monitor units (1,169 ± 395 vs. 1,338 ± 471, p < 0.001) and were simpler (MCSv 0.31 to 0.35; SAS10 mm 0.27 to 0.20; both p < 0.001).  Conclusion: The upgrade preserved quality-assurance compliance and produced simpler, lower-monitor-unit plans, but the systematic −1.5% offset between calculated and measured point dose persisted, showing the new MLC model alone does not remove it.
SN  - 0284-186X
SN  - 1651-226X
ER  -