Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy
Abstract
High-dose-rate (HDR) brachytherapy is highly conformal, but its accuracy depends on catheter and source placement, and clinical workflows lack real-time in vivo dosimetry (IVD) that verifies delivery without disrupting the procedure. We characterize a scintillator-based optical dosimeter integrated into a transrectal probe that tracks source position, speed, dose rate, and time-resolved cumulative dose. A 67 mm x 5 mm line scintillator is imaged by a CMOS camera through a 30-degree endoscopic lens; geometric and flat-field corrections remove barrel distortion and vignetting. Absolute dose was calibrated with a TOPAS Monte Carlo model of the probe's air-cavity geometry. The system was validated in a water phantom with an Ir-192 source and needles at radial offsets of 7, 10, and 15 mm. Response varied by ±2\% along the 67 mm scintillator and was linear over 5-21 cGy/s (R2 ≥ 0.995). The Monte Carlo model showed a 3.75\% dose-rate reduction relative to TG-43 water-equivalent conditions. In phantom, the system resolved a 3 mm needle displacement, confirmed by CT. Measured dose-rate profiles agreed with Monte Carlo in peak magnitude and spatial distribution, and integral dose agreed within 1\%. By mapping dose directly to the measurement plane of a standard ultrasound transducer, this system provides high-resolution, time-resolved verification of HDR delivery and a clinical tool for detecting delivery errors with minimal changes to existing treatment protocols.
Related
- Impact of tissue heterogeneity on intracranial Cesium-131 brachytherapy dosimetry: A patient-specific Monte Carlo Study.
- Experimental Investigation of Sub-aperture Spacing Effects on Scintillation Suppression in Multi-aperture Free-space Optical Communication under Atmospheric Turbulence
- The RUBIK satellite payload concept: a 3D-segmented plastic scintillation-based particle tracking detector
- Target-Dose Robustness During Major Pelvic Deformation After HDR Prostate Brachytherapy: Fraction-Level and Accumulated-Dose Effects of Online Adaptive Radiotherapy.
- Three-dimensional deep learning nomogram for post-plan quality assurance in high-dose-rate prostate brachytherapy.
BibTeX
@misc{willy2026optical,
title = {{Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy}},
author = {Kevin Willy and Megan Clark and Petr Bruza and Shauna McVorran and David Gladstone},
howpublished = {arXiv},
year = {2026},
month = nov,
doi = {10.48550/arxiv.2611.27146},
eprint = {2611.27146},
archivePrefix = {arXiv},
url = {https://arxiv.org/abs/2611.27146},
note = {CC BY 4.0},
}
RIS
TY - UNPB
TI - Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy
AU - Kevin Willy
AU - Megan Clark
AU - Petr Bruza
AU - Shauna McVorran
AU - David Gladstone
PY - 2026
DA - 2026/11/20
PB - arXiv
DO - 10.48550/arxiv.2611.27146
UR - https://arxiv.org/abs/2611.27146
AB - High-dose-rate (HDR) brachytherapy is highly conformal, but its accuracy depends on catheter and source placement, and clinical workflows lack real-time in vivo dosimetry (IVD) that verifies delivery without disrupting the procedure. We characterize a scintillator-based optical dosimeter integrated into a transrectal probe that tracks source position, speed, dose rate, and time-resolved cumulative dose. A 67 mm x 5 mm line scintillator is imaged by a CMOS camera through a 30-degree endoscopic lens; geometric and flat-field corrections remove barrel distortion and vignetting. Absolute dose was calibrated with a TOPAS Monte Carlo model of the probe's air-cavity geometry. The system was validated in a water phantom with an Ir-192 source and needles at radial offsets of 7, 10, and 15 mm. Response varied by ±2\% along the 67 mm scintillator and was linear over 5-21 cGy/s (R^2 ≥ 0.995). The Monte Carlo model showed a 3.75\% dose-rate reduction relative to TG-43 water-equivalent conditions. In phantom, the system resolved a 3 mm needle displacement, confirmed by CT. Measured dose-rate profiles agreed with Monte Carlo in peak magnitude and spatial distribution, and integral dose agreed within 1\%. By mapping dose directly to the measurement plane of a standard ultrasound transducer, this system provides high-resolution, time-resolved verification of HDR delivery and a clinical tool for detecting delivery errors with minimal changes to existing treatment protocols.
ER -