Standardised superior vena cava isolation with a balloon-in-basket pulsed-field ablation catheter: feasibility and acute safety of two procedural workflows in 53 consecutive patients
- 1Department of Cardiology, Medical University of Lublin, Lublin, Poland
- 2Wojewódzki Szpital Specjalistyczny w Białej Podlaskiej, Biała Podlaska, Poland
Abstract
Purpose: Pulsed-field ablation (PFA) is established for pulmonary vein isolation, but data on superior vena cava (SVC) isolation are limited. We assessed the feasibility and acute safety of SVC isolation with the Volt™ balloon-in-basket PFA catheter using two sequentially applied standardised workflows.
Methods: Retrospective single-centre series of 53 consecutive patients undergoing atrial fibrillation ablation with adjunctive SVC isolation. Workflow 1 (n = 28) used phrenic-nerve pacing and tailored energy; Workflow 2 (n = 25) used low-energy PFA only without phrenic pacing; both shared mapping-guided sinus-node sparing and a 20-second flow-preservation manoeuvre. A pre-specified within-platform sensitivity analysis controlled for a concurrent catheter iteration (20 January 2026).
Results: Acute SVC isolation was achieved in all 53 patients (100%, 95% CI 93.3–100%). No phrenic-nerve injury and no permanent atrioventricular block occurred (0/53); one Workflow 1 patient had transient junctional rhythm that resolved spontaneously. Workflow-attributable differences were confined to SVC-specific metrics and persisted within a single catheter-platform era: Workflow 2 used low-energy applications exclusively, required more applications (median 5 vs. 4; p = 0.010), and had a longer SVC isolation time (10.7 vs. 8.5 min; p = 0.006). Shorter procedure and fluoroscopy times in Workflow 2 did not persist within a single catheter-platform era (p = 0.94, p = 0.35). Median follow-up 84 days (data censored 1 July 2026).
Conclusions: SVC isolation with the Volt™ balloon-in-basket PFA catheter is feasible with favourable acute safety. A simplified low-energy-only protocol without phrenic-nerve pacing is a reasonable alternative to a phrenic-paced tailored-energy protocol. Durability and rhythm outcomes require dedicated follow-up study.
Graphical abstract:
Supplementary Information: The online version contains supplementary material available at https://doi.org/10.1007/s10840-026-02409-4.
Related
- Feasibility and procedural considerations of a helix-fixation leadless pacemaker implantation in patients with left ventricular assist devices.
- Feasibility and Procedural Considerations of Balloon‐Expandable TAVR in Patients With Prior Endovascular Aortic Stent Grafts
- Standardized perioperative terminology for emergency cervical cerclage: A framework for feasibility and procedural complexity assessment
- Fully Automated Multi-Chamber Strain Profiling in Consecutive Patients Undergoing Transthoracic Echocardiography: Feasibility, Functional Clustering, and Prognostic Implications
- Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility Study
BibTeX
@article{wojcik2026standardised,
title = {{Standardised superior vena cava isolation with a balloon-in-basket pulsed-field ablation catheter: feasibility and acute safety of two procedural workflows in 53 consecutive patients}},
author = {Maciej Wójcik and Paweł Błaszkiewicz and Andrzej Głowniak},
journal = {Journal of Interventional Cardiac Electrophysiology},
year = {2026},
month = aug,
doi = {10.1007/s10840-026-02409-4},
url = {https://doi.org/10.1007/s10840-026-02409-4},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Standardised superior vena cava isolation with a balloon-in-basket pulsed-field ablation catheter: feasibility and acute safety of two procedural workflows in 53 consecutive patients
AU - Maciej Wójcik
AU - Paweł Błaszkiewicz
AU - Andrzej Głowniak
PY - 2026
DA - 2026/08/05
JO - Journal of Interventional Cardiac Electrophysiology
DO - 10.1007/s10840-026-02409-4
UR - https://doi.org/10.1007/s10840-026-02409-4
AB - Purpose: Pulsed-field ablation (PFA) is established for pulmonary vein isolation, but data on superior vena cava (SVC) isolation are limited. We assessed the feasibility and acute safety of SVC isolation with the Volt™ balloon-in-basket PFA catheter using two sequentially applied standardised workflows. Methods: Retrospective single-centre series of 53 consecutive patients undergoing atrial fibrillation ablation with adjunctive SVC isolation. Workflow 1 (n = 28) used phrenic-nerve pacing and tailored energy; Workflow 2 (n = 25) used low-energy PFA only without phrenic pacing; both shared mapping-guided sinus-node sparing and a 20-second flow-preservation manoeuvre. A pre-specified within-platform sensitivity analysis controlled for a concurrent catheter iteration (20 January 2026). Results: Acute SVC isolation was achieved in all 53 patients (100%, 95% CI 93.3–100%). No phrenic-nerve injury and no permanent atrioventricular block occurred (0/53); one Workflow 1 patient had transient junctional rhythm that resolved spontaneously. Workflow-attributable differences were confined to SVC-specific metrics and persisted within a single catheter-platform era: Workflow 2 used low-energy applications exclusively, required more applications (median 5 vs. 4; p = 0.010), and had a longer SVC isolation time (10.7 vs. 8.5 min; p = 0.006). Shorter procedure and fluoroscopy times in Workflow 2 did not persist within a single catheter-platform era (p = 0.94, p = 0.35). Median follow-up 84 days (data censored 1 July 2026). Conclusions: SVC isolation with the Volt™ balloon-in-basket PFA catheter is feasible with favourable acute safety. A simplified low-energy-only protocol without phrenic-nerve pacing is a reasonable alternative to a phrenic-paced tailored-energy protocol. Durability and rhythm outcomes require dedicated follow-up study. Graphical abstract: Supplementary Information: The online version contains supplementary material available at https://doi.org/10.1007/s10840-026-02409-4.
SN - 1383-875X