Peer-reviewedCC BY-NC-ND 4.0Indian Pacing and Electrophysiology Journal

Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)

4 affiliations
  1. 1Department of Cardiology, Vakfıkebir State Hospital, Trabzon, Türkiye
  2. 2Department of Cardiology, University of Health Sciences Trabzon Kanuni Training and Research Hospital, Trabzon, Türkiye
  3. 3Department of Physiotherapy and Rehabilitation, Tarsus University Faculty of Health Sciences, Mersin, Türkiye
  4. 4Department of Cardiology, University of Health Sciences Trabzon Ahi Evren Thoracic and Cardiovascular Surgery Training and Research Hospital, Trabzon, Türkiye

Abstract

Background: Accurate assessment of volume status is essential in heart failure (HF). This study evaluated the association between cardiac implantable electronic device (CIED)-derived fluid monitoring and bioelectrical impedance analysis (BIA) for systemic volume assessment.

Methods: In this prospective cross-sectional study, 57 patients with HF carrying implantable cardioverter-defibrillators or cardiac resynchronization therapy devices with intrathoracic impedance-based fluid monitoring were enrolled. Patients were classified according to the presence of device-detected fluid alerts. Body composition and total body water (TBW) percentage were assessed by BIA (Tanita). Clinical, laboratory, echocardiographic, and device-derived data were compared. Logistic regression identified variables independently associated with fluid alerts.

Results: Patients with device-detected fluid alerts had higher TBW and lower body fat percentages than those without alerts (both p < 0.001). TBW showed a strong positive correlation with device-detected fluid alerts (r = 0.675, p < 0.001). A TBW threshold of 57.15% predicted fluid alerts with 80% sensitivity and 80% specificity (AUC 0.938, 95% CI 0.856–0.999; p < 0.001). Patients with fluid alerts also demonstrated more pronounced echocardiographic signs of right-sided congestion. In multivariable analysis, TBW remained the only independent predictor of device-detected fluid alerts (OR 1.33, 95% CI 1.07–1.66; p = 0.011).

Conclusion: BIA-derived TBW was strongly associated with device-detected fluid alerts and may serve as a practical complementary method for systemic volume assessment, particularly when device-based fluid monitoring is unavailable.

Related

  1. Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults
    Elodie Piche, Emeline Michel, Olivier Guerin +2
    SensorsPeer-reviewedCC BY 4.0Full textPhysiologyHAL
  2. Bioelectrical impedance analysis versus dual-energy X-ray absorptiometry for sarcopenia assessment in type 2 diabetes mellitus: A systematic review and meta-analysis
    Shobhit Raj, Sahana Shetty, Renuka Suvarna +2
    PLOS OnePeer-reviewedCC BY 4.0Full textGeneral biochemistryGeneral biologyPLOSPMC
  3. Comparative safety outcomes of three venous access routes for totally implantable venous access port placement and exploratory assessment of pre-removal catheter-related thrombosis in breast cancer patients
    Mintao Li, Chao Ma, Feng Lu +5
    Frontiers in oncologyPeer-reviewedCC BYFull textOncologyEurope PMCCrossref
  4. Bioelectrical Impedance Analysis–Derived Phase Angle in Initial Medical and Nursing Assessment for Cardiac Rehabilitation: Association with Cardiac Rehabilitation Risk Category and Longitudinal Change During the Program—An Exploratory Prospective Observational Study
    Francisco José Rodríguez-Velasco, José Alberto Becerra-Mejías, Luis Miguel Lara-Godoy +3
    Journal of Clinical MedicinePeer-reviewedCC BY 4.0General medicineCrossref
  5. Substituent effects on electronic properties and predicted acetylcholinesterase binding potential of donepezil analogues: a DFT, molecular docking, and ADMET study with dynamics assessment of the lead candidate
    Muhammad Mohsin Ejaz, Hamza Iftikhar, K. S. Al-Namshah +6
    In Silico PharmacologyPeer-reviewedLicence not statedFull textPharmacology & toxicologyPMC
BibTeX
@article{altıntepe2026comparison,
  title = {{Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)}},
  author = {Sedat Altıntepe and Fatih Gülçebi and Ahmet Özderya and Nurel Bellur and Ezgi Kalaycıoğlu},
  journal = {Indian Pacing and Electrophysiology Journal},
  year = {2026},
  month = aug,
  doi = {10.1016/j.ipej.2026.08.001},
  url = {https://doi.org/10.1016/j.ipej.2026.08.001},
  note = {CC BY-NC-ND 4.0},
}
RIS
TY  - JOUR
TI  - Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)
AU  - Sedat Altıntepe
AU  - Fatih Gülçebi
AU  - Ahmet Özderya
AU  - Nurel Bellur
AU  - Ezgi Kalaycıoğlu
PY  - 2026
DA  - 2026/08/03
JO  - Indian Pacing and Electrophysiology Journal
DO  - 10.1016/j.ipej.2026.08.001
UR  - https://doi.org/10.1016/j.ipej.2026.08.001
AB  - Background: Accurate assessment of volume status is essential in heart failure (HF). This study evaluated the association between cardiac implantable electronic device (CIED)-derived fluid monitoring and bioelectrical impedance analysis (BIA) for systemic volume assessment.  Methods: In this prospective cross-sectional study, 57 patients with HF carrying implantable cardioverter-defibrillators or cardiac resynchronization therapy devices with intrathoracic impedance-based fluid monitoring were enrolled. Patients were classified according to the presence of device-detected fluid alerts. Body composition and total body water (TBW) percentage were assessed by BIA (Tanita). Clinical, laboratory, echocardiographic, and device-derived data were compared. Logistic regression identified variables independently associated with fluid alerts.  Results: Patients with device-detected fluid alerts had higher TBW and lower body fat percentages than those without alerts (both p < 0.001). TBW showed a strong positive correlation with device-detected fluid alerts (r = 0.675, p < 0.001). A TBW threshold of 57.15% predicted fluid alerts with 80% sensitivity and 80% specificity (AUC 0.938, 95% CI 0.856–0.999; p < 0.001). Patients with fluid alerts also demonstrated more pronounced echocardiographic signs of right-sided congestion. In multivariable analysis, TBW remained the only independent predictor of device-detected fluid alerts (OR 1.33, 95% CI 1.07–1.66; p = 0.011).  Conclusion: BIA-derived TBW was strongly associated with device-detected fluid alerts and may serve as a practical complementary method for systemic volume assessment, particularly when device-based fluid monitoring is unavailable.