Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)
4 affiliations
- 1Department of Cardiology, Vakfıkebir State Hospital, Trabzon, Türkiye
- 2Department of Cardiology, University of Health Sciences Trabzon Kanuni Training and Research Hospital, Trabzon, Türkiye
- 3Department of Physiotherapy and Rehabilitation, Tarsus University Faculty of Health Sciences, Mersin, Türkiye
- 4Department of Cardiology, University of Health Sciences Trabzon Ahi Evren Thoracic and Cardiovascular Surgery Training and Research Hospital, Trabzon, Türkiye
Abstract
Background: Accurate assessment of volume status is essential in heart failure (HF). This study evaluated the association between cardiac implantable electronic device (CIED)-derived fluid monitoring and bioelectrical impedance analysis (BIA) for systemic volume assessment.
Methods: In this prospective cross-sectional study, 57 patients with HF carrying implantable cardioverter-defibrillators or cardiac resynchronization therapy devices with intrathoracic impedance-based fluid monitoring were enrolled. Patients were classified according to the presence of device-detected fluid alerts. Body composition and total body water (TBW) percentage were assessed by BIA (Tanita). Clinical, laboratory, echocardiographic, and device-derived data were compared. Logistic regression identified variables independently associated with fluid alerts.
Results: Patients with device-detected fluid alerts had higher TBW and lower body fat percentages than those without alerts (both p < 0.001). TBW showed a strong positive correlation with device-detected fluid alerts (r = 0.675, p < 0.001). A TBW threshold of 57.15% predicted fluid alerts with 80% sensitivity and 80% specificity (AUC 0.938, 95% CI 0.856–0.999; p < 0.001). Patients with fluid alerts also demonstrated more pronounced echocardiographic signs of right-sided congestion. In multivariable analysis, TBW remained the only independent predictor of device-detected fluid alerts (OR 1.33, 95% CI 1.07–1.66; p = 0.011).
Conclusion: BIA-derived TBW was strongly associated with device-detected fluid alerts and may serve as a practical complementary method for systemic volume assessment, particularly when device-based fluid monitoring is unavailable.
Related
- Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults
- Bioelectrical impedance analysis versus dual-energy X-ray absorptiometry for sarcopenia assessment in type 2 diabetes mellitus: A systematic review and meta-analysis
- Comparative safety outcomes of three venous access routes for totally implantable venous access port placement and exploratory assessment of pre-removal catheter-related thrombosis in breast cancer patients
- Bioelectrical Impedance Analysis–Derived Phase Angle in Initial Medical and Nursing Assessment for Cardiac Rehabilitation: Association with Cardiac Rehabilitation Risk Category and Longitudinal Change During the Program—An Exploratory Prospective Observational Study
- Substituent effects on electronic properties and predicted acetylcholinesterase binding potential of donepezil analogues: a DFT, molecular docking, and ADMET study with dynamics assessment of the lead candidate
BibTeX
@article{altıntepe2026comparison,
title = {{Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)}},
author = {Sedat Altıntepe and Fatih Gülçebi and Ahmet Özderya and Nurel Bellur and Ezgi Kalaycıoğlu},
journal = {Indian Pacing and Electrophysiology Journal},
year = {2026},
month = aug,
doi = {10.1016/j.ipej.2026.08.001},
url = {https://doi.org/10.1016/j.ipej.2026.08.001},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)
AU - Sedat Altıntepe
AU - Fatih Gülçebi
AU - Ahmet Özderya
AU - Nurel Bellur
AU - Ezgi Kalaycıoğlu
PY - 2026
DA - 2026/08/03
JO - Indian Pacing and Electrophysiology Journal
DO - 10.1016/j.ipej.2026.08.001
UR - https://doi.org/10.1016/j.ipej.2026.08.001
AB - Background: Accurate assessment of volume status is essential in heart failure (HF). This study evaluated the association between cardiac implantable electronic device (CIED)-derived fluid monitoring and bioelectrical impedance analysis (BIA) for systemic volume assessment. Methods: In this prospective cross-sectional study, 57 patients with HF carrying implantable cardioverter-defibrillators or cardiac resynchronization therapy devices with intrathoracic impedance-based fluid monitoring were enrolled. Patients were classified according to the presence of device-detected fluid alerts. Body composition and total body water (TBW) percentage were assessed by BIA (Tanita). Clinical, laboratory, echocardiographic, and device-derived data were compared. Logistic regression identified variables independently associated with fluid alerts. Results: Patients with device-detected fluid alerts had higher TBW and lower body fat percentages than those without alerts (both p < 0.001). TBW showed a strong positive correlation with device-detected fluid alerts (r = 0.675, p < 0.001). A TBW threshold of 57.15% predicted fluid alerts with 80% sensitivity and 80% specificity (AUC 0.938, 95% CI 0.856–0.999; p < 0.001). Patients with fluid alerts also demonstrated more pronounced echocardiographic signs of right-sided congestion. In multivariable analysis, TBW remained the only independent predictor of device-detected fluid alerts (OR 1.33, 95% CI 1.07–1.66; p = 0.011). Conclusion: BIA-derived TBW was strongly associated with device-detected fluid alerts and may serve as a practical complementary method for systemic volume assessment, particularly when device-based fluid monitoring is unavailable.
SN - 0972-6292
ER -