Experimental study of TPU and SEBS additively manufactured leaflets for novel aortic valve polymer TAVI prosthesis
6 affiliations
- 1Research Center for Genetics and Life Sciences, Sirius University of Science and Technology, Sirius, Russia
- 2Biofluids Laboratory, Perm National Research Polytechnic University, Perm, Russia
- 3Endovascular Surgery Department, E.N. Meshalkin National Medical Research Center, Novosibirsk, Russia
- 4Endovascular Surgery Department, Center of Surgery named after A. Vishnevsky, Moscow, Russia
- 5Department of Mechanical, Bioresources and Biomedical Engineering, School of Engineering, College of Science, Engineering and Technology, University of South Africa, Florida, South Africa
- 6Department of Mechanical Engineering, Faculty of Engineering and the Built Environment, Durban University of Technology, Durban, South Africa
Abstract
Background and Objective: Transcatheter valve implantation is a cutting-edge technology for high and low risk patients with aortic stenosis. Bioprosthetic valves are limited due to several disadvantages as a long-term durability and structural valve degeneration. Polymeric aortic valve leaflet devices represent state-of-the-art technology to overcome limitations of bioprostheses. Nevertheless, more recent biocompatible and biostable polymers have revealed a significant potential. Additionally, 3D additive manufacturing opens new paths for patient-specific design of biomedical implants. Flexible polymers may not only mimic native aortic valve, but also may improve long-term durability. Polyurethanes and styrene polymers are promising materials for modern polymer valves. Nevertheless, mechanical properties of additive manufactured TPU or SEBS leaflets are not properly studied.
Methods: Experimental study of TPU and SEBS samples and aortic valve leaflets for novel aortic valve polymer TAVI prosthesis is presented. Tensile tests and compression tests of three types of FDM-printed samples with various filament alignments (axial alignment, transversal alignment, diagonal alignment) were performed. The results of tests confirmed that longitudinal axial alignment achieve a maximum stress compared to transversal and diagonal directions. Compressive tests of TPU and SEBS 3D-manufactured aortic valve leaflets with digital image correlation technology were performed. Young’s modulus and Poisson’s ratio were evaluated. Moreover, parameters of constitutive models were found.
Results: The TPU tensile specimen exhibits the highest stress/strain values at a speed of 100 mm/min, regardless of the filament alignment pattern. Slightly lower results are observed at a speed of 50 mm/min, which is likely associated with a wide range of plastic deformation. In the case of SEBS, a different pattern is observed: at 500 mm/min, it demonstrates high resistance to fracture, unlike TPU. A comparison of TPU and SEBS specimens at the same speeds and with the same alignment, based on maximum stress values, shows a minimal difference of 4% with the axial alignment at a speed of 50 mm/min; however, when comparing strains, the difference increases to 90%. The results of the compression experiment on aortic valve specimens made of TPU and SEBS at a speed of 30 mm/min showed that TPU force values for is considerably lower (4% of the maximum case), while for SEBS the spread increases to 20% of the maximum case.
Conclusion: SEBS aortic valve leaflets were shown to be stiffer than TPU. It can be a step for choice of stiffer/softer leaflets for patient-specific device. The proposed approach can help manufacturers in elaboration and manufacturing of new TAVI polymer devices design made of TPU and SEBS polymers for clinical applications.
Related
- Surface Roughness as a Determinant of the Immunological Tissue Response to Additively Manufactured Titanium Implants
- Effect of Surface Roughness and Polyelectrolyte Multilayer Coatings on the In Vitro Cytocompatibility of Additively Manufactured Titanium Implants
- Effect of localized thermal history on martensitic microstructure and microhardness variation in wire-arc additively manufactured 17–4 PH stainless steel beads
- Impact of aortic valve leaflets calcium volume and distribution on Post-TAVR conduction abnormalities
- Developing and evaluating patient information leaflets for artificial intelligence use in healthcare
BibTeX
@article{kuchumov2026experimental,
title = {{Experimental study of TPU and SEBS additively manufactured leaflets for novel aortic valve polymer TAVI prosthesis}},
author = {Alex G. Kuchumov and Denis Khoroshev and Aleksandra Kutuzova and Aleksei Baranov and Oleg Krestyaninov and Ekaterina Barbashina and Pavel Zelenchuk and Edward Matjeka and Thanyani Pandelani and Fulufhelo Nemavhola},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2026},
month = oct,
doi = {10.3389/fbioe.2026.1959014},
url = {https://doi.org/10.3389/fbioe.2026.1959014},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Experimental study of TPU and SEBS additively manufactured leaflets for novel aortic valve polymer TAVI prosthesis
AU - Alex G. Kuchumov
AU - Denis Khoroshev
AU - Aleksandra Kutuzova
AU - Aleksei Baranov
AU - Oleg Krestyaninov
AU - Ekaterina Barbashina
AU - Pavel Zelenchuk
AU - Edward Matjeka
AU - Thanyani Pandelani
AU - Fulufhelo Nemavhola
PY - 2026
DA - 2026/10/21
JO - Frontiers in Bioengineering and Biotechnology
DO - 10.3389/fbioe.2026.1959014
UR - https://doi.org/10.3389/fbioe.2026.1959014
AB - Background and Objective: Transcatheter valve implantation is a cutting-edge technology for high and low risk patients with aortic stenosis. Bioprosthetic valves are limited due to several disadvantages as a long-term durability and structural valve degeneration. Polymeric aortic valve leaflet devices represent state-of-the-art technology to overcome limitations of bioprostheses. Nevertheless, more recent biocompatible and biostable polymers have revealed a significant potential. Additionally, 3D additive manufacturing opens new paths for patient-specific design of biomedical implants. Flexible polymers may not only mimic native aortic valve, but also may improve long-term durability. Polyurethanes and styrene polymers are promising materials for modern polymer valves. Nevertheless, mechanical properties of additive manufactured TPU or SEBS leaflets are not properly studied. Methods: Experimental study of TPU and SEBS samples and aortic valve leaflets for novel aortic valve polymer TAVI prosthesis is presented. Tensile tests and compression tests of three types of FDM-printed samples with various filament alignments (axial alignment, transversal alignment, diagonal alignment) were performed. The results of tests confirmed that longitudinal axial alignment achieve a maximum stress compared to transversal and diagonal directions. Compressive tests of TPU and SEBS 3D-manufactured aortic valve leaflets with digital image correlation technology were performed. Young’s modulus and Poisson’s ratio were evaluated. Moreover, parameters of constitutive models were found. Results: The TPU tensile specimen exhibits the highest stress/strain values at a speed of 100 mm/min, regardless of the filament alignment pattern. Slightly lower results are observed at a speed of 50 mm/min, which is likely associated with a wide range of plastic deformation. In the case of SEBS, a different pattern is observed: at 500 mm/min, it demonstrates high resistance to fracture, unlike TPU. A comparison of TPU and SEBS specimens at the same speeds and with the same alignment, based on maximum stress values, shows a minimal difference of 4% with the axial alignment at a speed of 50 mm/min; however, when comparing strains, the difference increases to 90%. The results of the compression experiment on aortic valve specimens made of TPU and SEBS at a speed of 30 mm/min showed that TPU force values for is considerably lower (4% of the maximum case), while for SEBS the spread increases to 20% of the maximum case. Conclusion: SEBS aortic valve leaflets were shown to be stiffer than TPU. It can be a step for choice of stiffer/softer leaflets for patient-specific device. The proposed approach can help manufacturers in elaboration and manufacturing of new TAVI polymer devices design made of TPU and SEBS polymers for clinical applications.
SN - 2296-4185
ER -