Comparison of the Melbourne rapid fields-web device-independent online perimeter and the octopus 600 perimeter in glaucoma patients

5 affiliations
  1. 1The University of Melbourne, Parkville, VIC, Australia
  2. 2Centre for Eye Research Australia, East Melbourne, VIC, Australia
  3. 3Royal Victorian Eye and Ear Hospital, East Melbourne, VIC, Australia
  4. 4Marengo Asia Hospitals, Gurugram, Haryana, India
  5. 5Department of Ophthalmology, Mayo Clinic, Jacksonville, FL, United States

Abstract

Introduction: The purpose of this study is to evaluate the agreement, repeatability, and testing efficiency of Melbourne Rapid Fields-Web (MRF-Web), a browser-based, device-independent perimeter, compared with the Octopus 600 perimeter in glaucoma patients.

Methods: This retrospective, cross-sectional, single-center observational study included 113 patients undergoing routine visual field assessment for glaucoma or glaucoma suspicion at Marengo Asia Hospitals, Gurgaon, India. MRF-Web testing was performed on a calibrated 14-inch laptop with Hindi voice guidance. Agreement between MRF-Web and Octopus 600 for Mean Deviation (MD) and pattern-based global indices was assessed using intraclass correlation coefficients (ICCs) and Bland–Altman analysis. Test-retest repeatability was evaluated in 51 patients who underwent repeat MRF-Web testing.

Results: MRF-Web showed strong group-level concordance with Octopus 600 for MD, with an ICC of 0.88 (95% CI, 0.81–0.92), Bland–Altman bias of -0.78 dB, and 95% limits of agreement from -4.72 to 3.17 dB. Concordance for the pattern-based indices was moderate, with an ICC of 0.72 (95% CI, 0.58–0.81), bias of -0.59 dB, and 95% limits of agreement from -4.41 to 3.24 dB. MRF-Web demonstrated good test-retest concordance, with an ICC of 0.86 (95% CI, 0.69–0.94). Mean per-eye test duration was significantly shorter with MRF-Web than with Octopus 600 (232.9 ± 49.8 versus 342.4 ± 94.4 seconds).

Conclusions: MRF-Web demonstrated strong group-level concordance with Octopus 600 MD measurements and shorter per-eye testing times in this predominantly mild-disease cohort. The limits of agreement indicate that measurements from the two platforms should not be regarded as directly interchangeable, and further study is required in moderate and advanced glaucoma.

Related

  1. Evaluation of Bruch's membrane opening minimum rim width and retinal nerve fiber layer in patients with glaucoma and comparison with normal population.
    Mustafa Davas, Halit Oguz
    BMC ophthalmologyPeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMCPMC
  2. Macular Retinal Ganglion Cell Counts and Central Visual Field Sensitivity in Glaucoma: A Comparison of Logarithmic and Linear Scales
    GIAMPAOLO TROIANO, ARIANNA ASADI, STEFANO PIO +3
    In VivoPeer-reviewedCC BY-NC-ND 4.0Full textGeneral medicinePMC
  3. Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer.
    Xiaoyuan Shen, Zidong Chen, Kaixin Tan +6
    Clinical ophthalmology (Auckland, N.Z.)Peer-reviewedCC BY-NCFull textGeneral medicineEurope PMCPMC
  4. Long-Term Outcomes and Safety of Non-Cyclodialytic, Bleb-Independent Surgical Enhancement of Uveoscleral Outflow in Open-Angle Glaucoma, with Evaluation of the Role of Conjunctival Lymphatic Vessels in Suprachoroidal Drainage.
    Vinod Kumar, Mikhail Aleksandrovich Frolov, Galina Nikolaevna Dushina +3
    Vision (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. Association between the stress hyperglycemia ratio and neovascular glaucoma in patients with proliferative diabetic retinopathy
    Chenwei Zhao, Jiahao Xu, Hongmei Qian +1
    Frontiers in endocrinologyPeer-reviewedCC BYFull textEndocrinology & metabolismEurope PMC
BibTeX
@article{kong2026comparison,
  title = {{Comparison of the Melbourne rapid fields-web device-independent online perimeter and the octopus 600 perimeter in glaucoma patients}},
  author = {George Y. X. Kong and Algis Vingrys and Manisha Ambawat and Md Mahboob and Shakira Bano and Syril Dorairaj and Shibal Bhartiya},
  journal = {Frontiers in Ophthalmology},
  year = {2026},
  month = oct,
  doi = {10.3389/fopht.2026.1961375},
  url = {https://doi.org/10.3389/fopht.2026.1961375},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Comparison of the Melbourne rapid fields-web device-independent online perimeter and the octopus 600 perimeter in glaucoma patients
AU  - George Y. X. Kong
AU  - Algis Vingrys
AU  - Manisha Ambawat
AU  - Md Mahboob
AU  - Shakira Bano
AU  - Syril Dorairaj
AU  - Shibal Bhartiya
PY  - 2026
DA  - 2026/10/12
JO  - Frontiers in Ophthalmology
DO  - 10.3389/fopht.2026.1961375
UR  - https://doi.org/10.3389/fopht.2026.1961375
AB  - Introduction: The purpose of this study is to evaluate the agreement, repeatability, and testing efficiency of Melbourne Rapid Fields-Web (MRF-Web), a browser-based, device-independent perimeter, compared with the Octopus 600 perimeter in glaucoma patients.  Methods: This retrospective, cross-sectional, single-center observational study included 113 patients undergoing routine visual field assessment for glaucoma or glaucoma suspicion at Marengo Asia Hospitals, Gurgaon, India. MRF-Web testing was performed on a calibrated 14-inch laptop with Hindi voice guidance. Agreement between MRF-Web and Octopus 600 for Mean Deviation (MD) and pattern-based global indices was assessed using intraclass correlation coefficients (ICCs) and Bland–Altman analysis. Test-retest repeatability was evaluated in 51 patients who underwent repeat MRF-Web testing.  Results: MRF-Web showed strong group-level concordance with Octopus 600 for MD, with an ICC of 0.88 (95% CI, 0.81–0.92), Bland–Altman bias of -0.78 dB, and 95% limits of agreement from -4.72 to 3.17 dB. Concordance for the pattern-based indices was moderate, with an ICC of 0.72 (95% CI, 0.58–0.81), bias of -0.59 dB, and 95% limits of agreement from -4.41 to 3.24 dB. MRF-Web demonstrated good test-retest concordance, with an ICC of 0.86 (95% CI, 0.69–0.94). Mean per-eye test duration was significantly shorter with MRF-Web than with Octopus 600 (232.9 ± 49.8 versus 342.4 ± 94.4 seconds).  Conclusions: MRF-Web demonstrated strong group-level concordance with Octopus 600 MD measurements and shorter per-eye testing times in this predominantly mild-disease cohort. The limits of agreement indicate that measurements from the two platforms should not be regarded as directly interchangeable, and further study is required in moderate and advanced glaucoma.