Transpulmonary Atrial Pacemaker Implantation in Fontan Circulation With Bilateral Glenn
4 affiliations
- 1Division of Pediatric Cardiology, Clínica Cardio VID, Universidad Pontificia Bolivariana, Medellín, Colombia
- 2Division of Cardiac Electrophysiology, Clínica Cardio VID, Medellín, Colombia
- 3Division of Pediatric Cardiology, Adult Congenital Heart Disease Program, Clínica Cardio VID, Medellín, Colombia
- 4Division of Cardiology, Echocardiography, Adult Congenital Heart Disease Program, Clínica Cardio VID, Medellín, Colombia
Abstract
Background: Patients with Fontan circulation frequently develop atrial arrhythmias and sinus node dysfunction. Atrial pacing is challenging because conventional transvenous access is limited by complex Fontan anatomy.
Case Summary: A 20-year-old man with dextroisomerism, bilateral Glenn anastomoses, and Fontan circulation presented with symptomatic sinus node dysfunction and recurrent supraventricular tachycardia. A transpulmonary transvenous approach achieved endocardial atrial lead implantation with excellent pacing parameters. Despite optimized medical therapy, heart failure progressed, and the patient died during transplant evaluation.
Discussion: This case highlights the technical feasibility of transpulmonary endocardial atrial pacing as an alternative to repeat epicardial surgery in selected patients with Fontan circulation. Careful procedural planning, multimodality imaging, and individualized assessment of long-term anticoagulation are important considerations. Despite arrhythmia control, optimized therapy and restoration of atrioventricular synchrony may not prevent heart failure progression, which may be irreversible.
Take-Home Messages: Transpulmonary endocardial atrial pacing is a feasible alternative in carefully selected Patients with Fontan circulation. Sudden cardiac death risk stratification remains limited in this study group.
Related
- Direct-visualization implantation of an atrial leadless pacemaker at tricuspid valve surgery.
- Leadless Pacemaker Implantation in Fabry Disease With Permanent Atrial Fibrillation and Bradycardia: Diagnostic and Implantation Guidance by Preprocedural Cardiac Magnetic Resonance Imaging.
- Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation
- Successful atrial leadless pacemaker implantation in an 11-year-old with congenital sinus node dysfunction: a case report
- Device-detected new-onset atrial fibrillation after dual-chamber pacemaker implantation in patients with sick sinus syndrome under managed ventricular pacing: a risk nomogram from the 3-month landmark
BibTeX
@article{ruales2026transpulmonary,
title = {{Transpulmonary Atrial Pacemaker Implantation in Fontan Circulation With Bilateral Glenn}},
author = {Andrea Dávila Ruales and Juan David Ramírez Barrera and Juan Fernando Agudelo Uribe and Margarita Maria Zapata Sánchez and Luz Adriana Ocampo Aristizábal and Rafael Correa Velásquez},
journal = {JACC Case Reports},
year = {2026},
month = nov,
doi = {10.1016/j.jaccas.2026.110858},
url = {https://doi.org/10.1016/j.jaccas.2026.110858},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Transpulmonary Atrial Pacemaker Implantation in Fontan Circulation With Bilateral Glenn
AU - Andrea Dávila Ruales
AU - Juan David Ramírez Barrera
AU - Juan Fernando Agudelo Uribe
AU - Margarita Maria Zapata Sánchez
AU - Luz Adriana Ocampo Aristizábal
AU - Rafael Correa Velásquez
PY - 2026
DA - 2026/11/11
JO - JACC Case Reports
DO - 10.1016/j.jaccas.2026.110858
UR - https://doi.org/10.1016/j.jaccas.2026.110858
AB - Background: Patients with Fontan circulation frequently develop atrial arrhythmias and sinus node dysfunction. Atrial pacing is challenging because conventional transvenous access is limited by complex Fontan anatomy. Case Summary: A 20-year-old man with dextroisomerism, bilateral Glenn anastomoses, and Fontan circulation presented with symptomatic sinus node dysfunction and recurrent supraventricular tachycardia. A transpulmonary transvenous approach achieved endocardial atrial lead implantation with excellent pacing parameters. Despite optimized medical therapy, heart failure progressed, and the patient died during transplant evaluation. Discussion: This case highlights the technical feasibility of transpulmonary endocardial atrial pacing as an alternative to repeat epicardial surgery in selected patients with Fontan circulation. Careful procedural planning, multimodality imaging, and individualized assessment of long-term anticoagulation are important considerations. Despite arrhythmia control, optimized therapy and restoration of atrioventricular synchrony may not prevent heart failure progression, which may be irreversible. Take-Home Messages: Transpulmonary endocardial atrial pacing is a feasible alternative in carefully selected Patients with Fontan circulation. Sudden cardiac death risk stratification remains limited in this study group.
SN - 2666-0849
ER -