Investigation of the dosimetric characteristics of real-time silicone/Bi2O3 nanocomposites for the diagnostic level X-rays
Abstract
This study presents the development of high-sensitivity real-time X-ray dosimeters based on room-temperature-vulcanized (RTV-2) silicone nanocomposites incorporating 0–60 wt% Bi ₂ O ₃ nanoparticles, fabricated via solution casting. Structural and morphological characterization using X-ray diffraction (XRD), scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX) with elemental mapping, and elemental analysis confirmed the uniform dispersion of monoclinic Bi ₂ O ₃ nanorods within the amorphous RTV-2 matrix. Complementary Fourier-transform infrared spectroscopy (FTIR), Raman spectroscopy, and thermogravimetric analysis (TGA) validated the chemical integrity, thermal stability, and mechanical robustness of the resulting nanocomposites. Dosimetric performance was investigated under diagnostic-range X-ray exposure using a COMET MXR 320/26 source (dose rates: 2.8–12 mGy/min). The I-V characteristics at 4.8 mGy/min exhibited a highly linear response over ± 400 V. At 400 V, the photocurrent increased from ~ 0.56 nA (pure RTV-2) to ~ 1.66 nA (60 wt% Bi ₂ O ₃ ), corresponding to a ~ 3-fold enhancement in photocurrent. The devices demonstrated good short-term repeatability (standard deviation of 2.3% over ten consecutive measurements) and a signal-to-noise ratio (SNR) of up to ~ 1900 at 400 V. Compared with reported values for commercial MOSFET dosimeters (~ 0.1–1 nA/mGy) and previously reported Bi ₂ O ₃ -polymer systems, the developed nanocomposites exhibit a higher competitive normalized sensitivity (~ 0.35 nA/(mGy/min), or 878 nC·mGy⁻ ¹ ·cm⁻ ³ on a volumetric basis), demonstrating their suitability as lead-free alternatives for low-dose, real-time diagnostic X-ray dosimetry; further evaluation of long-term stability and mechanical durability is needed to establish clinical applicability.
Related
- Microbeam radiotherapy - clinical and dosimetric investigation of a first clinical application.
- Dosimetric impact of baseline shift plan in a magnetic resonance linear accelerator according to target motion and plan characteristics.
- A combined ND diagnostic investigation by DHSPI, SIRT, NMR, THZ, on Giotto fresco
- Dosimetric and Treatment Delivery Efficiency Comparison of Two Ring-gantry Linear Accelerator Platforms for Liver Stereotactic Body Radiotherapy
- Dosimetric Comparison of Functional Radiosurgery Treatments for Trigeminal Neuralgia, Essential Tremor, and Obsessive-compulsive Disorder Using Gamma Knife, CyberKnife, and Linear Accelerators
BibTeX
@article{keyvani2026investigation,
title = {{Investigation of the dosimetric characteristics of real-time silicone/Bi2O3 nanocomposites for the diagnostic level X-rays}},
author = {Farideh Keyvani and Leila Gholamzadeh and Shahryar Malekie and Sedigheh Kashian},
journal = {Scientific Reports},
year = {2026},
month = oct,
doi = {10.1038/s41598-026-76522-9},
url = {https://www.nature.com/articles/s41598-026-76522-9},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Investigation of the dosimetric characteristics of real-time silicone/Bi2O3 nanocomposites for the diagnostic level X-rays
AU - Farideh Keyvani
AU - Leila Gholamzadeh
AU - Shahryar Malekie
AU - Sedigheh Kashian
PY - 2026
DA - 2026/10/18
JO - Scientific Reports
DO - 10.1038/s41598-026-76522-9
UR - https://www.nature.com/articles/s41598-026-76522-9
AB - This study presents the development of high-sensitivity real-time X-ray dosimeters based on room-temperature-vulcanized (RTV-2) silicone nanocomposites incorporating 0–60 wt% Bi ₂ O ₃ nanoparticles, fabricated via solution casting. Structural and morphological characterization using X-ray diffraction (XRD), scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX) with elemental mapping, and elemental analysis confirmed the uniform dispersion of monoclinic Bi ₂ O ₃ nanorods within the amorphous RTV-2 matrix. Complementary Fourier-transform infrared spectroscopy (FTIR), Raman spectroscopy, and thermogravimetric analysis (TGA) validated the chemical integrity, thermal stability, and mechanical robustness of the resulting nanocomposites. Dosimetric performance was investigated under diagnostic-range X-ray exposure using a COMET MXR 320/26 source (dose rates: 2.8–12 mGy/min). The I-V characteristics at 4.8 mGy/min exhibited a highly linear response over ± 400 V. At 400 V, the photocurrent increased from ~ 0.56 nA (pure RTV-2) to ~ 1.66 nA (60 wt% Bi ₂ O ₃ ), corresponding to a ~ 3-fold enhancement in photocurrent. The devices demonstrated good short-term repeatability (standard deviation of 2.3% over ten consecutive measurements) and a signal-to-noise ratio (SNR) of up to ~ 1900 at 400 V. Compared with reported values for commercial MOSFET dosimeters (~ 0.1–1 nA/mGy) and previously reported Bi ₂ O ₃ -polymer systems, the developed nanocomposites exhibit a higher competitive normalized sensitivity (~ 0.35 nA/(mGy/min), or 878 nC·mGy⁻ ¹ ·cm⁻ ³ on a volumetric basis), demonstrating their suitability as lead-free alternatives for low-dose, real-time diagnostic X-ray dosimetry; further evaluation of long-term stability and mechanical durability is needed to establish clinical applicability.
SN - 2045-2322
ER -