Peer-reviewedCC BY 4.0Hypertension Research

Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients

  1. 1Department of Geriatric and General Medicine, Graduate School of Medicine, The University of Osaka, Suita, Japan
  2. 2Division of Public Health, Department of Social Medicine, Graduate School of Medicine, The University of Osaka, Suita, Japan
  3. 3Arblet Inc., Tokyo, Japan

Abstract

Accurate ambulatory blood pressure monitoring (ABPM) is essential for cardiovascular risk assessment and hypertension management, but conventional cuff-based ABPM causes discomfort and sleep disturbance, potentially affecting measurement reliability. The European Society of Hypertension (ESH) established evaluation criteria for cuffless blood pressure devices, with the awake/asleep test as the primary evaluation test for automated wearable devices. This study evaluated whether Alysis-001, a software-based cuffless blood pressure estimation algorithm using sensor data from a wrist-worn device, meets ESH awake/asleep test criteria in hypertensive patients under two data extraction strategies: (1) arm-stabilized periods only ( ≥ 10 s static); (2) including low-intensity activity (high-intensity excluded). We prospectively enrolled 92 hypertensive outpatients receiving antihypertensive treatment who simultaneously wore Alysis Wear and underwent conventional 24-h ABPM; 78 completed the study. Under both strategies, Alysis-001 demonstrated excellent agreement with reference ABPM across all measurement periods, with intraclass correlation coefficients exceeding 0.70 for systolic and diastolic blood pressure. Alysis-001 successfully met all ESH awake/asleep test criteria under both strategies: mean difference between awake-asleep blood pressure changes ≤5 mmHg, standard deviation ≤8 mmHg, and correlation coefficients ≥0.70 for both systolic and diastolic blood pressure. Alysis-001 also demonstrated high sensitivity, specificity, and accuracy in identifying non-dippers under both strategies. To our knowledge, this represents the first evaluation of a Device Type 3 cuffless blood pressure algorithm (automated, wearable, demographic-calibrated, not at heart level) meeting ESH awake/asleep test criteria, suggesting Alysis-001’s potential as a clinically viable alternative for ambulatory blood pressure monitoring in hypertensive patients.

Related

  1. Clinicopathological Review of Renovascular Hypertension and Hypertensive Emergency-related Nephropathy Based on Case Series
    Yoshifumi Ubara, Kei Kono, Naoki Sawa
    Internal MedicinePeer-reviewedCC BY-NC-NDFull textGeneral medicineJ-STAGE
  2. Hypertension in People with Multiple Sclerosis: Epidemiology, Mechanisms, Hypertensive Disorders of Pregnancy, and Clinical Implications.
    Ahmed Arafa, Amira S A Said, Ehab Elkady +3
    Diseases (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  3. Assessment of Persistent Postpartum Hypertension Following Hypertensive Disorders of Pregnancy: A Scoping Review
    Marzieh Daneshmand, Harika Dasari, Lindsay Hales +7
    Journal of the American Heart AssociationPeer-reviewedCC BY-NC-NDFull textObstetrics, gynaecology & paediatricsCrossrefDOAJ
  4. Knowledge of hypertension and understanding of home blood pressure monitoring among hypertensive patients in India: results from the GRAND Study
    Anuj Maheshwari, Noriko Matsushita, Rituparna Barooah +20
    Hypertension ResearchPeer-reviewedCC BY 4.0Full text1 citedCardiovascular medicinePMC
  5. Prevalence, clinical features and emergency department blood pressure management of hypertensive emergencies and uncontrolled hypertension: a retrospective multi-centre study
    Alessandro Maloberti, Giulia Capitoli, Tommaso Valobra +17
    Hypertension ResearchPeer-reviewedCC BY 4.0Full textCardiovascular medicinePMC
BibTeX
@article{hongyo2026evaluation,
  title = {{Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients}},
  author = {Kazuhiro Hongyo and Atsushi Hirayama and Kosuke Shimizu and Taisuke Ueno and Taku Fujimoto and Akitoshi Hara and Hiroshi Koriyama and Yoichi Takami and Yoshio Iwashima and Ryo Kawasaki and Koichi Yamamoto},
  journal = {Hypertension Research},
  year = {2026},
  month = aug,
  doi = {10.1038/s41440-026-02728-z},
  url = {https://doi.org/10.1038/s41440-026-02728-z},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients
AU  - Kazuhiro Hongyo
AU  - Atsushi Hirayama
AU  - Kosuke Shimizu
AU  - Taisuke Ueno
AU  - Taku Fujimoto
AU  - Akitoshi Hara
AU  - Hiroshi Koriyama
AU  - Yoichi Takami
AU  - Yoshio Iwashima
AU  - Ryo Kawasaki
AU  - Koichi Yamamoto
PY  - 2026
DA  - 2026/08/21
JO  - Hypertension Research
DO  - 10.1038/s41440-026-02728-z
UR  - https://doi.org/10.1038/s41440-026-02728-z
AB  - Accurate ambulatory blood pressure monitoring (ABPM) is essential for cardiovascular risk assessment and hypertension management, but conventional cuff-based ABPM causes discomfort and sleep disturbance, potentially affecting measurement reliability. The European Society of Hypertension (ESH) established evaluation criteria for cuffless blood pressure devices, with the awake/asleep test as the primary evaluation test for automated wearable devices. This study evaluated whether Alysis-001, a software-based cuffless blood pressure estimation algorithm using sensor data from a wrist-worn device, meets ESH awake/asleep test criteria in hypertensive patients under two data extraction strategies: (1) arm-stabilized periods only ( ≥ 10 s static); (2) including low-intensity activity (high-intensity excluded). We prospectively enrolled 92 hypertensive outpatients receiving antihypertensive treatment who simultaneously wore Alysis Wear and underwent conventional 24-h ABPM; 78 completed the study. Under both strategies, Alysis-001 demonstrated excellent agreement with reference ABPM across all measurement periods, with intraclass correlation coefficients exceeding 0.70 for systolic and diastolic blood pressure. Alysis-001 successfully met all ESH awake/asleep test criteria under both strategies: mean difference between awake-asleep blood pressure changes ≤5 mmHg, standard deviation ≤8 mmHg, and correlation coefficients ≥0.70 for both systolic and diastolic blood pressure. Alysis-001 also demonstrated high sensitivity, specificity, and accuracy in identifying non-dippers under both strategies. To our knowledge, this represents the first evaluation of a Device Type 3 cuffless blood pressure algorithm (automated, wearable, demographic-calibrated, not at heart level) meeting ESH awake/asleep test criteria, suggesting Alysis-001’s potential as a clinically viable alternative for ambulatory blood pressure monitoring in hypertensive patients.
SN  - 0916-9636
SN  - 1348-4214
ER  -