Mobile radiation shielding device for scatter dose reduction during portable radiography in an orthopedic ward
- 1Department of Biomedical Sciences and Biomedical Engineering, Faculty of Medicine, Prince of Songkla University, Hat Yai, Songkhla, Thailand
- 2Radiological Technology Program, Department of Radiology, Faculty of Medicine, Prince of Songkla University, Hat Yai, Songkhla, Thailand
Abstract
Portable radiography in orthopedic wards exposes healthcare professionals and adjacent patients to scatter radiation in open, shared clinical environments. In crowded wards, maintaining an optimal distance or clearing the area during exposure is often impractical. This study designed, constructed, and evaluated a foldable mobile radiation shielding system comprising two three-panel units, each with a chromed steel frame (210 × 180 cm) with 0.6-mm nominal lead sheets covered by a vinyl outer layer (66 × 100 cm) per panel, mounted on wheeled castors for bedside deployment. The two units were deployed together, one along each lateral side of the bed. Scatter dose rates were measured using an radiation survey meter and a portable X-ray unit during phantom radiography across common orthopedic projections at eight angular positions (0°–315° in 45° increments) and five distances (50–250 cm). The two-unit configuration achieved a median scatter dose reduction of 94.1% (IQR: 88.0%–97.7%; Wilcoxon W = 51,360, N = 320, p < 0.001). Reductions were broadly consistent across tube voltages (65–85 kVp) and remained substantial across all measurement distances (84.3%–89.6% across 50–250 cm), but varied significantly by angular position, with the highest reduction at 90° (mean 96.91%) and the lowest reductions recorded at 0° (mean 54.02%) and 180° (mean 75.69%). These two directions represented the open axial ends of the setup and were not directly covered by the side-mounted shielding panels. Under a representative high-workload scenario, the two-unit configuration reduced estimated annual occupational doses across all projections, with all post-shielding means remaining well below the average annual occupational dose limit of 20 mSv/year. User feedback was more favorable among radiologic technologists (mean 4.80/5) than among orthopedic ward nurses (mean 3.67/5). The two-unit configuration showed substantial scatter reduction and may serve as an additional protective measure during portable radiography in orthopedic ward settings, with positioning guidance required to maximize protection at clinically relevant angular locations.
Related
- Body Size, Radiation Dose and Exposure Index in Conventional Radiography: A Multicentre Prospective Study
- Can we finally shed the “lead”? Initial Polish experience with the RAMPART Defender radiation shielding system
- Sustainable Diagnostics: A Comparative Analysis of Radiation, Carbon Emissions, and the Cost of Weightbearing CT, Radiography, and Standard CT in Weber B Ankle Fracture Imaging; a Short Report
- Evaluation of Neuroangiography Settings and Protective Shielding on Radiation Exposure to Patient and Interventionist: A Comparative Dosimetric Study
- Development and validation of a pregnant phantom with adjustable fetal presentation for region-specific fetal radiation dose assessment during pelvic radiography.
BibTeX
@article{cheewakul2026mobile,
title = {{Mobile radiation shielding device for scatter dose reduction during portable radiography in an orthopedic ward}},
author = {Jongwat Cheewakul and Natee Ina and Suphalak Khamruang Marshall and Sitthichok Chaichulee},
journal = {PLOS One},
year = {2026},
month = nov,
doi = {10.1371/journal.pone.0362759},
url = {https://doi.org/10.1371/journal.pone.0362759},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Mobile radiation shielding device for scatter dose reduction during portable radiography in an orthopedic ward
AU - Jongwat Cheewakul
AU - Natee Ina
AU - Suphalak Khamruang Marshall
AU - Sitthichok Chaichulee
PY - 2026
DA - 2026/11/13
JO - PLOS One
DO - 10.1371/journal.pone.0362759
UR - https://doi.org/10.1371/journal.pone.0362759
AB - Portable radiography in orthopedic wards exposes healthcare professionals and adjacent patients to scatter radiation in open, shared clinical environments. In crowded wards, maintaining an optimal distance or clearing the area during exposure is often impractical. This study designed, constructed, and evaluated a foldable mobile radiation shielding system comprising two three-panel units, each with a chromed steel frame (210 × 180 cm) with 0.6-mm nominal lead sheets covered by a vinyl outer layer (66 × 100 cm) per panel, mounted on wheeled castors for bedside deployment. The two units were deployed together, one along each lateral side of the bed. Scatter dose rates were measured using an radiation survey meter and a portable X-ray unit during phantom radiography across common orthopedic projections at eight angular positions (0°–315° in 45° increments) and five distances (50–250 cm). The two-unit configuration achieved a median scatter dose reduction of 94.1% (IQR: 88.0%–97.7%; Wilcoxon W = 51,360, N = 320, p < 0.001). Reductions were broadly consistent across tube voltages (65–85 kVp) and remained substantial across all measurement distances (84.3%–89.6% across 50–250 cm), but varied significantly by angular position, with the highest reduction at 90° (mean 96.91%) and the lowest reductions recorded at 0° (mean 54.02%) and 180° (mean 75.69%). These two directions represented the open axial ends of the setup and were not directly covered by the side-mounted shielding panels. Under a representative high-workload scenario, the two-unit configuration reduced estimated annual occupational doses across all projections, with all post-shielding means remaining well below the average annual occupational dose limit of 20 mSv/year. User feedback was more favorable among radiologic technologists (mean 4.80/5) than among orthopedic ward nurses (mean 3.67/5). The two-unit configuration showed substantial scatter reduction and may serve as an additional protective measure during portable radiography in orthopedic ward settings, with positioning guidance required to maximize protection at clinically relevant angular locations.
SN - 1932-6203
ER -