Ultra‐Sensitive Amperometric Detection of Ciprofloxacin in Environmental and Biological Matrices Using an ITO Modified With Tungsten Trioxide‐Based Nanocomposite
Abstract
Although significant progress has been achieved in the electrochemical sensing of ciprofloxacin (CIP), there remains a critical need for highly sensitive and reliable platforms capable of rapid charge transfer and accurate detection at trace levels in complex environmental and biological matrices. In this work, a tungsten trioxide (WO 3 )/titanium carbide (TiC)/graphitic carbon nitride (g‐C 3 N 4 ) ternary nanocomposite was synthesized via a facile hydrothermal method and structurally characterized by XRD, XPS, FTIR, UV–vis, PL and SEM. UV–vis analysis revealed a reduced band gap for the nanocomposite compared to the individual components, indicating an improved electronic structure and enhanced light absorption. Furthermore, photoluminescence (PL) studies demonstrated a significant reduction in emission intensity, indicating effective charge separation and suppressed electron–hole recombination. Electrochemical investigations demonstrated a significantly enhanced response towards CIP in comparison to the individual components. Cyclic voltammetry (CV) showed a higher oxidation current, while electrochemical impedance spectroscopy (EIS) confirmed a reduced charge transfer resistance, evidencing strong synergistic interactions within the ternary system. This enhanced performance is attributable to the complementary functions of TiC, which improves electrical conductivity, WO 3 , which provides numerous electroactive sites, and g‐C 3 N 4 , which facilitates charge transfer and analyte adsorption. The developed sensor demonstrated a broad linear range from 10 to 200 µM ‘CIP’, an ultra‐low detection limit (LOD) of 0.02 µM and an exceptional sensitivity of 7.97 µA µM −1 cm −2 . Furthermore, the sensor's performance has been demonstrated by the excellent selectivity and satisfactory recoveries in real samples, including river water, tap water and saliva. This demonstrates its practical applicability for environmental monitoring.
Related
- Fe2O3–MgO–CuO@MWCNT Nanocomposite for Highly Sensitive Electrochemical Detection of l -Cysteine at Low Potential in Biological Samples
- MnO2–Co3O4@RGO hybrid nanocomposite-modified carbon paste electrode for sensitive electrochemical detection of hydrogen peroxide in food and edible-oil matrices
- Plasmonic Quantum Nanocomposite-Modified Multimode–Multicore Fiber for Sensitive Optical Biosensing of Triclosan
- A new La-complex immobilized on GO Ni-MNC nanocomposite as a green, returnable, robust and homoselective nanocatalyst in the synthesis of biological organic compounds
- Green synthesis and theoretical study of novel biological active naphthyridine via multicomponent reactions employing ZnO/SiO2/Fe3O4 as effective magnetic nanocomposite
BibTeX
@article{nahdi2026ultrasensitive,
title = {{Ultra‐Sensitive Amperometric Detection of Ciprofloxacin in Environmental and Biological Matrices Using an ITO Modified With Tungsten Trioxide‐Based Nanocomposite}},
author = {Amira Nahdi and Mabrouka Ghiloufi and Fathi Touati and Hassouna Dhaouadi and Tobias Schnabel and Nicole Jaffrezic-Renault and Salah Kouass and Hamdi Ben Halima},
journal = {Electrochemical Science Advances},
year = {2026},
month = jul,
doi = {10.1002/elsa.70032},
url = {https://hal.science/hal-05798424v1},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Ultra‐Sensitive Amperometric Detection of Ciprofloxacin in Environmental and Biological Matrices Using an ITO Modified With Tungsten Trioxide‐Based Nanocomposite
AU - Amira Nahdi
AU - Mabrouka Ghiloufi
AU - Fathi Touati
AU - Hassouna Dhaouadi
AU - Tobias Schnabel
AU - Nicole Jaffrezic-Renault
AU - Salah Kouass
AU - Hamdi Ben Halima
PY - 2026
DA - 2026/07/01
JO - Electrochemical Science Advances
DO - 10.1002/elsa.70032
UR - https://hal.science/hal-05798424v1
AB - Although significant progress has been achieved in the electrochemical sensing of ciprofloxacin (CIP), there remains a critical need for highly sensitive and reliable platforms capable of rapid charge transfer and accurate detection at trace levels in complex environmental and biological matrices. In this work, a tungsten trioxide (WO 3 )/titanium carbide (TiC)/graphitic carbon nitride (g‐C 3 N 4 ) ternary nanocomposite was synthesized via a facile hydrothermal method and structurally characterized by XRD, XPS, FTIR, UV–vis, PL and SEM. UV–vis analysis revealed a reduced band gap for the nanocomposite compared to the individual components, indicating an improved electronic structure and enhanced light absorption. Furthermore, photoluminescence (PL) studies demonstrated a significant reduction in emission intensity, indicating effective charge separation and suppressed electron–hole recombination. Electrochemical investigations demonstrated a significantly enhanced response towards CIP in comparison to the individual components. Cyclic voltammetry (CV) showed a higher oxidation current, while electrochemical impedance spectroscopy (EIS) confirmed a reduced charge transfer resistance, evidencing strong synergistic interactions within the ternary system. This enhanced performance is attributable to the complementary functions of TiC, which improves electrical conductivity, WO 3 , which provides numerous electroactive sites, and g‐C 3 N 4 , which facilitates charge transfer and analyte adsorption. The developed sensor demonstrated a broad linear range from 10 to 200 µM ‘CIP’, an ultra‐low detection limit (LOD) of 0.02 µM and an exceptional sensitivity of 7.97 µA µM −1 cm −2 . Furthermore, the sensor's performance has been demonstrated by the excellent selectivity and satisfactory recoveries in real samples, including river water, tap water and saliva. This demonstrates its practical applicability for environmental monitoring.
SN - 2698-5977