Effect of radiotherapy on cardiac implantable electronic devices - evaluation of potential explanatory variables in more than a decade.
4 affiliations
- 1Dept. of Radiation Oncology, Hannover Medical School, Carl- Neuberg-Str. 1, 30625, Hannover, Germany.
- 2Department of Cardiology and Angiology, Hannover Medical School, Hannover Heart Rhythm Center, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
- 3Department of Cardiology and Angiology, Hannover Medical School, Hannover Heart Rhythm Center, Carl-Neuberg-Str. 1, 30625, Hannover, Germany. [email protected].
- 4Dept. of Radiation Oncology, Hannover Medical School, Carl- Neuberg-Str. 1, 30625, Hannover, Germany. [email protected].
Abstract
Background and purpose
As the population ages, an increasing number of patients are undergoing radiotherapy with an implanted cardiac device. Therefore, ensuring the safety of these devices during radiotherapy is crucial. This study aims to analyse the effect of radiotherapy on such devices, with a particular focus on changes in lead variables.Methods
This retrospective study examined all patients with a cardiac implantable electronic device (CIED) who underwent radiotherapy at a single institution between 2012 and 2022. Device function and associated changes in lead parameters were analysed before and after radiotherapy.Results
The study cohort included 279 patients who underwent 310 courses of radiotherapy. During the 11-year study period, one device failure requiring replacement due to a software error was recorded. This occurred following irradiation with 15 MV photons. Changes in individual lead parameters were more common, occurring in almost half of all radiotherapy courses. Lead impedances were particularly sensitive to changes, with high-voltage impedances being affected more frequently (61% of impedances) than other impedances. Sensing, pacing thresholds, and battery voltage were less frequently affected, with changes occurring in fewer than 10% of the analysed leads or batteries. None of the observed changes manifested as clinical symptoms. There was no statistically significant association between radiation parameters and the observed changes in lead values with only a few courses using neutron-generating photon energies. Furthermore, no dose-effect relationship for device doses could be established. Receiving multiple courses of radiotherapy in one year or over several years did not increase the risk to the device or cause greater variation in lead variables.Conclusion
Only one safety event occurred during the observation period, resulting in an error rate of 0.4% across 11 years and 279 patients. Changes in individual lead parameters were more frequent. However, no clear association could be established between the observed changes and radiation parameters. In particular, changes in lead variables were not explained by the dose to the device or repeated courses of radiotherapy. Therefore, the risk to the safety and function of cardiac devices from radiotherapy appears to be small.
Related
- Increase in Reimbursement Does Not Necessarily Place a Financial Burden on Patients ― Background Behind the Reimbursement Revision for Remote Monitoring of Cardiac Implantable Electronic Devices in Japan ―
- Changes in Remote Monitoring for Patients With Cardiac Implantable Electronic Devices After a Reimbursement Revision ― An Interrupted Time-Series Analysis ―
- Cutaneous warning signs of cardiac implantable electronic device infection: a pictorial guide for clinicians
- Comparison of device-based and bioelectrical impedance methods for volume status assessment in heart failure patients with cardiac implantable electronic devices (CIED vs. BIA fluid assessment in heart failure)
- Enacting policy to use administrative data for surveillance of cardiac implantable electronic device infections in Alberta, Canada
BibTeX
@article{christiansen2026effect,
title = {{Effect of radiotherapy on cardiac implantable electronic devices - evaluation of potential explanatory variables in more than a decade.}},
author = {Alexander Christiansen and Johanna Müller-Leisse and Henrike A K Hillmann and Hans Christiansen and David Duncker and Roland Merten},
journal = {Radiation oncology (London, England)},
year = {2026},
month = nov,
doi = {10.1186/s13014-026-02940-8},
url = {https://europepmc.org/article/MED/43053142},
note = {CC BY},
}
RIS
TY - JOUR
TI - Effect of radiotherapy on cardiac implantable electronic devices - evaluation of potential explanatory variables in more than a decade.
AU - Alexander Christiansen
AU - Johanna Müller-Leisse
AU - Henrike A K Hillmann
AU - Hans Christiansen
AU - David Duncker
AU - Roland Merten
PY - 2026
DA - 2026/11/18
JO - Radiation oncology (London, England)
DO - 10.1186/s13014-026-02940-8
UR - https://europepmc.org/article/MED/43053142
AB - Background and purpose As the population ages, an increasing number of patients are undergoing radiotherapy with an implanted cardiac device. Therefore, ensuring the safety of these devices during radiotherapy is crucial. This study aims to analyse the effect of radiotherapy on such devices, with a particular focus on changes in lead variables.Methods This retrospective study examined all patients with a cardiac implantable electronic device (CIED) who underwent radiotherapy at a single institution between 2012 and 2022. Device function and associated changes in lead parameters were analysed before and after radiotherapy.Results The study cohort included 279 patients who underwent 310 courses of radiotherapy. During the 11-year study period, one device failure requiring replacement due to a software error was recorded. This occurred following irradiation with 15 MV photons. Changes in individual lead parameters were more common, occurring in almost half of all radiotherapy courses. Lead impedances were particularly sensitive to changes, with high-voltage impedances being affected more frequently (61% of impedances) than other impedances. Sensing, pacing thresholds, and battery voltage were less frequently affected, with changes occurring in fewer than 10% of the analysed leads or batteries. None of the observed changes manifested as clinical symptoms. There was no statistically significant association between radiation parameters and the observed changes in lead values with only a few courses using neutron-generating photon energies. Furthermore, no dose-effect relationship for device doses could be established. Receiving multiple courses of radiotherapy in one year or over several years did not increase the risk to the device or cause greater variation in lead variables.Conclusion Only one safety event occurred during the observation period, resulting in an error rate of 0.4% across 11 years and 279 patients. Changes in individual lead parameters were more frequent. However, no clear association could be established between the observed changes and radiation parameters. In particular, changes in lead variables were not explained by the dose to the device or repeated courses of radiotherapy. Therefore, the risk to the safety and function of cardiac devices from radiotherapy appears to be small.
SN - 1748-717X
ER -