Loss of Biventricular Capture After CRT-D Implantation
- 1Department of Cardiology, Henry Ford Health Providence, Southfield, Michigan, USA
- 2Department of Internal Medicine, Henry Ford Health Providence, Southfield, Michigan, USA
Abstract
Case Summary: A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction.
Take-Home Message: Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.
Related
- Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation
- Biventricular Reverse Remodeling After Percutaneous Pulmonary Valve Implantation With a Self-Expandable Pulsta Valve in Native Right Ventricular Outflow Tract
- Residual Hearing Loss After Chronic Cochlear Implantation in Guinea Pigs is Primarily Conductive
- Improvement in health-related quality of life after complex biventricular repair.
- Beyond Strut Coverage ― Toward Functional Endothelization After Drug-Eluting Stent Implantation ―
BibTeX
@article{weber2026loss,
title = {{Loss of Biventricular Capture After CRT-D Implantation}},
author = {Paul Weber and Dhruva Govil and Dipak Shah},
journal = {JACC Case Reports},
year = {2026},
month = sep,
doi = {10.1016/j.jaccas.2026.110182},
url = {https://doi.org/10.1016/j.jaccas.2026.110182},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Loss of Biventricular Capture After CRT-D Implantation
AU - Paul Weber
AU - Dhruva Govil
AU - Dipak Shah
PY - 2026
DA - 2026/09/02
JO - JACC Case Reports
DO - 10.1016/j.jaccas.2026.110182
UR - https://doi.org/10.1016/j.jaccas.2026.110182
AB - Case Summary: A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction. Take-Home Message: Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.
SN - 2666-0849
ER -