Loss of Biventricular Capture After CRT-D Implantation

  1. 1Department of Cardiology, Henry Ford Health Providence, Southfield, Michigan, USA
  2. 2Department of Internal Medicine, Henry Ford Health Providence, Southfield, Michigan, USA

Abstract

Case Summary: A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction.

Take-Home Message: Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.

Related

  1. Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation
    Matthew A. Soderstrom, Ariana Ramirez, John Henry Dean +3
    JACC Case ReportsPeer-reviewedCC BY-NC-ND 4.0Full textCardiovascular medicinePMC
  2. Biventricular Reverse Remodeling After Percutaneous Pulmonary Valve Implantation With a Self-Expandable Pulsta Valve in Native Right Ventricular Outflow Tract
    Woo Young Park, Gi Beom Kim, Sang Yun Lee +5
    Circulation JournalPeer-reviewedLicence not statedFull textCardiovascular medicineJ-STAGE
  3. Residual Hearing Loss After Chronic Cochlear Implantation in Guinea Pigs is Primarily Conductive
    Wenxuan He, Rubing Xing, Jordan D. Hill +3
    bioRxivPreprintCC BYFull textPhysiologybioRxivPMC
  4. Improvement in health-related quality of life after complex biventricular repair.
    Kathryn O Stack, Kimberlee Gauvreau, Sarah Campos +5
    JTCVS openPeer-reviewedCC BY-NC-NDFull textEpidemiology & public healthEurope PMCPMC
  5. Beyond Strut Coverage ― Toward Functional Endothelization After Drug-Eluting Stent Implantation ―
    Takanori Yamazaki, Keisuke Kojima
    Circulation JournalPeer-reviewedCC BY-NC-NDFull textSurgery & anaesthesiaJ-STAGE
BibTeX
@article{weber2026loss,
  title = {{Loss of Biventricular Capture After CRT-D Implantation}},
  author = {Paul Weber and Dhruva Govil and Dipak Shah},
  journal = {JACC Case Reports},
  year = {2026},
  month = sep,
  doi = {10.1016/j.jaccas.2026.110182},
  url = {https://doi.org/10.1016/j.jaccas.2026.110182},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Loss of Biventricular Capture After CRT-D Implantation
AU  - Paul Weber
AU  - Dhruva Govil
AU  - Dipak Shah
PY  - 2026
DA  - 2026/09/02
JO  - JACC Case Reports
DO  - 10.1016/j.jaccas.2026.110182
UR  - https://doi.org/10.1016/j.jaccas.2026.110182
AB  - Case Summary: A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction.  Take-Home Message: Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.
SN  - 2666-0849
ER  -